When the Well-plate Handler is switched ON, the Well-plate Handler arm moves to
WA R N I N G
the vertical and horizontal home positions. Ensure that you are out of the way of the
arm before switching ON the Well-plate Handler.
8 Verify correct fuse replacement by switching ON the Well-plate Handler

Moving the Well-plate Handler

1 Remove any well-plates from the Input and Output racks.
2 Switch OFF the power and unplug the power cord from the power outlet
3 Prepare a space to accommodate the Well-plate Handler during the move
The Well-plate Handler weighs approximately 32 pounds (14.53 kg) and should be
WA R N I N G
lifted with care. Take the proper precautions to avoid injury.
4 Lift the Well-plate Handler straight up and off the locating pins of the
5 Disconnect the locating plate screws and remove the Well-plate Handler
Do not lose any of the screws or spacers.
C A U T I O N
6 To reinstall the Well-plate Handler, follow the instructions in
Agilent 1200 Series Sample Capacity Extension User Manual
power - the Well-plate Handler should initialize immediately.
and from the receptacle on the rear panel of the instrument. Disconnect the
communications cable from the rear panel of the instrument.
(an empty area on the lab bench or a sturdy cart).
Well-plate Handler Locating Plate and set it carefully on the bench or cart.
Depending on the distance that you are moving the instrument, you may
wish to repackage the Well-plate Handler in its original shipping carton.
Locating Plate, setting it aside with the Well-plate Handler.
"Installation"" to assemble and re-calibrate the Well-plate Handler.