Agilent Technologies 1200 Series User Manual page 31

Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

Modifying the Well-plate Handler for 10-inch Racks
If you intend to use the Well-plate Handler with 10-inch racks instead of
8.5-inch racks, you need to make the following modifications:
1 Unscrew the button extension from the underside of the gripper arm (see
Figure 19
You can move the gripper arm while the Well-plate Handler is switched off without
N O T E
damaging the instrument.
2 Remove the two Phillips-head screws from the gripper fingers at each side
3 Install the new gripper finger set that was delivered with the 10 inch rack
Agilent 1200 Series Sample Capacity Extension User Manual
Figure
19).
The button extension (left) and removed (right)
of the gripper, and remove the fingers.
kit, with the fingers pointing down. Ensure that all screws are tight.