Installing a new automation interface

If a defective automation interface has been replaced:
1 Disconnect all connections to the Automation Interface
2 Remove the two lock screws holding the Automation Interface (see
3 Withdraw the Automation Interface from the high performance
4 Install the new automation interface as described
1 Turn on the Agilent 1200 Series LC system including the high performance
2 Start the ChemStation SW. Switch to the Diagnosis view and select
3 Type the serial number of the new automation interface and reset the high
52
Figure
10)
autosampler and replace the lock screws in position in the Automation
Interface.
Automation Interface"
autosampler. The autosampler will abort the initialization and the status
LED will turn red.
Maintenance > High Performance ALS > New Automation Interface.
performance autosampler.
on page 13.