Installation
6 Insert the voltage selector card into the power entry port housing so that
7 Loosen the Phillips head screw on the fuse pedestal two turns. Remove the
Fuse
Figure 16
8 Invert the fuse block so the two 1.0 Amp fuses are on top for 230–240V
9 Confirm that the fuses are seated securely in the fuse block
10 Slide the fuse block back into position and tighten the screw. Make sure that
22
The indicator pin is correctly positioned for 230V. The pin tip will be visible
at the 230 "window" in the cover.
the selected voltage points inward and the indicator pin faces outwards
(towards the cover).
fuse block by sliding it up and away from the screw and pedestal, as shown
in
Figure
16.
Removing the fuse-block
operation, or the single 1.5 Amp fuse is on top for 110–120V operation.
the active fuse(s) face(s) out, see
Figure
17.