Adjusting For The Gripper Fingers - Agilent Technologies 1200 Series User Manual

Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

6 Adjust the position of the Well-plate Handler until the plate is lowered
7 Tighten the screws.

Adjusting for the gripper fingers

We recommend that you make these adjustments using the 54-vial tray. This is especially
N O T E
important if you intend to use the 54-vial tray in your work.
1 Place a plate on the transport position of the Automation Interface.
2 Use the Test functions of the Well-plate Handler to
3 If the plate moves when you close the gripper fingers, loosen the three
4 Adjust the position of the Well-plate Handler until there is no movement of
5 Tighten the screws.
6 Loosen the four screws holding the gripper fingers.
7 Use the Test functions of the Well-plate Handler to close the gripper fingers.
8 Tighten the four screws.
Agilent 1200 Series Sample Capacity Extension User Manual
b loosen the three screws on the adjustment plate that allow you to move
the Well-plate Handler sideways relative to the Automation Interface.
exactly on to the transport position of the Automation Interface.
a position the gripper arm above the Automation Interface (R6),
b move the arm down until it encounters the plate.
c close and open the gripper fingers.
screws on the adjustment plate that allow you to move the Well-plate
Handler forward and back relative to the Automation Interface.
the plate when you close the gripper fingers.
This ensures that the gripper fingers are correctly aligned with the plate.
Installation
2