Installation
2
Installing the Optional Fixed Barcode Reader
The fixed barcode reader can be installed on either the side of the Automation
Interface (the portrait position), or in one of three positions at the rear (the
landscape position). Its position, and how you place the plates in the racks, will
be determined by where you place the barcodes on the plates (see
Figure 24
and
Figure
25).
Barcode reader in portrait position
Barcode
Plate location A1
Rack
Well-plate
Automation
Extension
Handler
Interface
Figure 24
Barcodes on the ends of the plates
The default position is the portrait position. With this arrangement, you can read 60-mm
N O T E
long barcodes with the hand-held barcode reader when the plates are mounted in the rack.
If you want to use the landscape position, and also read the plates with the hand-held
barcode reader when they are in the rack, the barcodes must be no longer than 40 mm.
Agilent 1200 Series Sample Capacity Extension User Manual
29