Adjusting the Lateral Position of the Well-plate Handler

The position of the Well-plate Handler must be aligned with the Automation
Interface so that
• the transport position of the Automation Interface is at the same radius as
• the well-plate in the transport position of the Automation Interface is
The adjustment is made in three stages:
1 A preliminary adjustment, described in
2 Well-plate handler teaching, described in
3 A fine adjustment, described in
38
the racks,
located centrally between the gripper fingers.
Well-plate Handler Position"
and gripper are moved manually.
page 41, in which the rotary and vertical positions of the well-plate handler
gripper arm are calibrated.
Position"
on page 48, using the Test functions of the Agilent ChemStation or
Hand-held Controller to align the Well-plate Handler with the Automation
Interface.
"Preliminary Adjustment of the
on page 39, where the Well-plate Handler arm
"Well-plate Handler Teaching"
"Fine Adjustment of the Well-plate Handler
Agilent 1200 Series Sample Capacity Extension User Manual