Barcode Requirements - Agilent Technologies 1200 Series User Manual

Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

If you want to change the configuration of the hand-held barcode reader, for
example to read Code 39 or Code 2 of 5 interleaved with checksum, or
Codabar with transfer of start/stop, you must read each of the seven barcodes
in
Barcodes 5, 6 and 7 have two options. You select only one each of these options to install.
N O T E
If, at a later date, you want to change any of these options, you must reinstall all seven
commands in the correct sequence.
1 To read a barcode, point the barcode reader at the barcode and press the

Barcode Requirements

For reliable and accurate reading of barcodes, observe the following
requirements:
• Line thickness must be between 0.19 and 1 mm.
• Maximum width of the barcode is 60 mm.
• A clear protecting foil is allowed.
• For end-mounted barcodes, the top edge of the bars must be 9 mm or more
Use the Barcode Reader Test function (see
page 89) to check the readability of your barcodes.
Agilent 1200 Series Sample Capacity Extension User Manual
Table 1
on page 34, in the sequence shown in the table.
trigger. An audible note confirms that the barcode has been registered.
above the bottom edge of the plate.
"Automation Interface Test"
Installation
2
on
33

Advertisement

Table of Contents
loading

Table of Contents