Installation
Store the button extension and the gripper fingers for the 8.5-inch racks carefully in a
C A U T I O N
safe place, in case you need to convert the Well-plate handler back to 8.5-inch rack
configuration.
Installing the Input and Output Racks
Rack support disk
Rack locating pin
Rack locating disk
Figure 20
1 Locate the hole for the Rack Locating Pin in the bottom of the rack.
2 Place the rack on the white Rack Locating Disk so the Rack Locating Pin fits
3 Place the other rack on the Well-plate Handler in the same manner (see
26
Input Rack Location
into the hole on the bottom of the rack. The racks should be in portrait
orientation (short edge of the well-plate closest to the Well-plate Handler)
with the channel for the Well-plate Handler arm towards the center of the
Well-plate Handler.
Figure
21).
Agilent 1200 Series Sample Capacity Extension User Manual