Using the ChemStation
Using the Hand-held Controller
3 Adjust the height of the barcode reader until the two LEDs display green
Figure 31
4 When the two LEDs display green lights, tighten the two screws at either
5 Turn OFF the Barcode Adjustment mode in the Agilent ChemStation or the
6 Use the Test Barcode Reader function of the Agilent ChemStation or the
Agilent 1200 Series Sample Capacity Extension User Manual
a Switch to the Diagnostics view, and display the Automation Interface Test
dialog box (see
"Testing with the ChemStation"
b Click the Barcode Adjustment button. The barcode reader reads
continuously.
a From the Tests screen, display the Automation Interface Tests.
b Select Barcode Adjustment On and click Execute. The barcode reader reads
continuously.
lights. You can view the status of the LEDs using the mirror plate below the
barcode reader mounting plate (see
The mirror plate below the barcode reader mounting plate
side of the barcode reader mounting plate.
Hand-held Controller.
Hand-held Controller to check the accuracy and consistency of the barcode
reading.
on page 89).
Figure
31).