Agilent Technologies 1200 Series User Manual page 49

Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

Move in plate
Eject plate
Rotary Teaching Instructions
1 Click Set Default Values.
2 Select position R1.
3 Click Move Down until the arm sits just above the rack.
4 Click Plus 10 and/or Plus to move the arm clockwise, or Minus 10 and/or
Figure 33
5 Select a new rack position and repeat instructions 3 and 4.
All positions must be taught, including those of the optional rack extension, if fitted.
N O T E
6 Select position R6 (Automation Interface).
7 Click Move Down until the arm sits just above the transport position.
8 Repeat instructions 3 and 4.
9 Click Save to save the coordinates.
Agilent 1200 Series Sample Capacity Extension User Manual
Minus to move the arm counter-clockwise until the gripper is centralized in
the rack, as in
Figure
Gripper arm centralized in the rack
Moves the plate into the autosampler
Ejects the plate from the autosampler
33.