2
Installation
Changing the Fuse Configuration, if necessary
The Well-plate Handler is delivered for 230–240V operation using two 1.0 Amp
fuses. To convert the unit from the dual 1.0 Amp fuse configuration used for
230-240V operation to the single 1.5 Amp fuse configuration used for 100-120V
operation, refer to the diagrams and instructions below. Otherwise, proceed to
"Seating the Well-plate Handler on its Locating Plate"
Electrical shock hazard.
WA R N I N G
Disconnect the power cord before changing the fuses.
For continued fire protection and correct functioning of the unit, replace fuses only
with fuses of the same type and rating.
1 Ensure the Well-plate Handler's power switch is in the OFF position and
2 Using a small flat-blade screwdriver, gently pry the cover/fuse block
Figure 13
3 Grasp the voltage selector card (not the indicator pin which is attached to
20
the power cord is unplugged.
assembly away from the power entry port housing and remove (see
Figure
13).
Prying open the cover/fuse-block assembly
it) using needle-nosed pliers and remove it from the housing (see
Figure
14).
Agilent 1200 Series Sample Capacity Extension User Manual
on page 23.