Adjusting Fixture Press Travel on the Press Down Model (U9401A)

During testing, inadequate travel of the press may result from loose
contacts while excessive travel could damage the fixture. To correct the
travel distance, adjust the position of the lower position sensor.
To correct excessive travel, move the lower position sensor up. To correct
inadequate travel, move the sensor down.
Figure 43 Rear view of Press Down Unit showing sensors and sensor switch
(a)
1
2
3
4
5
32
Open the access panel at the top of the Press Down Unit.
Disable the sensors by flipping the sensor switch down (a).
Press both DOWN buttons to lower the press to the correct position.
Move the lower position sensor (b) till the sensor light turns on. Secure
the sensor.
Reactivate the sensors by flipping the sensor switch up.
(b)