4.2.3 Factory Defaults

Parameter
Vent Mode
TV (tidal volume)
Pinsp
RR
Mech RR
Tinsp
I:E
Trig. Window
Flow Trig.
End Breath
Psupport
PEEP
Tpause
Backup Time
O2%
Balance Gas
Circuit
Gas Outlet installed
Paw Color
Flow Color
Resp Color
CO2 Color
Paw Units
Weight Units
Temperature Units
CO2 Units
Altitude
Gas Supply Pressure
Units
Decimal marker
Language
Gas Supply Colors
O2 flow tube
Vent drive gas
PSV Pro
SIMV/PSV
PCV
SIMV-PC
N2O enabled
1009-0357-000 11/03
The following table lists the factory defaults for parameters and alarm limits:
Value
VCV
500 ml
5 cmH2O
(5 hPa, 0.5 kPa, 5 mbar, 3.6
mmHg)
12 /min
12 /min
1.70
1:2.0
25%
2 l/min
25%
Off
Off
Off
30 s
100%
Air
Circle
SCGO
Yellow
Green
White
White
cmH2O
kg
C
%
300 m
kPa
. <dot>
English
ISO
(O2 = white, N2O = blue,
Air = blk/wht)
Right side
O2
Enabled
Enabled
Enabled
Enabled
No
4 Installation and Service Menus
Alarm Limit
Value
Pmax High
40 cmH2O
(40 hPa, 4 kPa, 40 mbar,
30 mmHg)
MV High
10 l/min
MV Low
2 l/min
TV High
1000 ml
TV Low
Off
RR High
Off
RR Low
Off
Et CO2 High
6.5%
(50 mmHg or 6.5kPa)
Et CO2 Low
Off
Fi CO2 High
Off
Fi O2High
Off
Fi O2 Low
21%
EtO2 High
Off
EtO2 Low
Off
Fi Iso High
5%
Fi Iso Low
Off
Et Iso High
Off
Et Iso Low
Off
Fi Sev High
8%
Fi Sev Low
Off
Et Sev High
Off
Et Sev Low
Off
Fi Des High
15%
Fi Des Low
Off
Et Des High
Off
Et Des Low
Off
Fi Enf High
5%
Fi Enf Low
Off
Et Enf High
Off
Et Enf Low
Off
Fi Hal High
5%
FI Hal Low
Off
Et Hal High
Off
Et Hal Low
Off
4-5