7.4 Steps and Messages displayed during the System Checkout

7.4.1 Steps for the
Quick Check
1009-0357-000 11/03
Step1: Check to make sure bag/vent switch is set correctly.
• If switch is set to manual mode continue with the next step.
• If set to ventilator mode fail with, "Wrong circuit selected".
Step2: Is O
available and working?
2
• If O
Supply is Not low and mixer passes a mixer flow test of 250 ml/min of
2
O
continue with the next step.
2
• If not fail with "Low O
Step3: Is circuit (airway) pressure too high?
• Check the circuit (airway) pressure is less than 30 cm H
next step.
• If airway pressure greater than 30 cm H
• Step4: Find manual circuit leak:
• Increase flow to find the manual circuit leak.
• If pressure does not increase to 30 cm H
circuit".
• If flow required to maintain pressure at 30 cm H
state "Manual circuit leak is XXX ml/min at 30 cm H
• If flow required to maintain pressure at 30 cm H
250 cm H
0 continue with the next step.
2
Step 5: Run mixer tests on the O
test):
• If O
pressure is low, fail with "Low O
2
• If mixer does not fail the 3L,10L and the leak test, continue with the next step.
• If mixer fails the 3L,10L or the leak test, fail with the mixer failure.
Step 6: Run mixer tests on the Air channel (check supply and run a 3L, 10L and leak
test):
• If O
is drive gas and air supply is low, continue with next step and indicate,
2
"Could not test air".
• If Air is the drive gas and air supply is low, continue with next step and
indicate, "Ventilator has not drive gas".
• If mixer does not fail the 3L,10L and the leak test, continue with the next step.
If mixer fails the 3L,10L or the leak test, fail with the mixer failure.
supply pressure" or with mixer failure.
2
0 fail with "Circuit pressure too high".
2
0 fail with "Cannot pressurize
2
2
channel (check supply and run a 3L, 10L and leak
2
Supply pressure".
2
7 Troubleshooting
0; continue with the
2
0 is greater than 250 ml/min
0".
2
0 is less than or equal to
2
7-45