Datex-Ohmeda S/5 Avance Technical Reference Manual page 164

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance
Error Log Entry
Alarm Text
Action/Troubleshooting
MGAS LINE BLOCKED
Sample line blocked
>20 SEC
Replace sample line. See AM TRM for further Troubleshooting.
MGAS SAMPLE LINE
Check D-Fend
NOT CONNECTED
>40 S
Replace D-Fend. See AM TRM for further Troubleshooting.
MGAS SENSOR INOP
Module fail. No CO
> XX
AA, O
See AM TRM for further Troubleshooting.
MIXER BAL GAS
Alternate O
CHANGE FAIL
Reboot System. If problem continues, replace the Mixer.
MIXER BALGAS Flow
FAIL
Reboot System. If problem continues, replace the Mixer.
MIXER O2 Flow FAIL
Alternate O
Reboot System. If problem continues, replace the Mixer.
MIXER ACB COM FAIL
Alternate O
Reboot System. If problem continues, replace the Mixer.
7-36
Condition (Basic info)
MGAS SPEC states The sample
tubing inside or outside the
monitor blocked, or the water
trap is occluded.
MGAS SPEC states The sample
tubing or the D-Fend module is
not installed
,
MGAS SPEC Mgas
2
communicates hardware failure
data
2
(RAM failure; ROM checksum
error; Error in CPU eeprom; Error
O
preamp eeprom; Error in SSS
2
board eeprom; Voltage error;
Lamp control failure.) or UPI does
not initialize
Screen
Mixer Status Bit:
2
STS__BALGAS_CHANGE_OVER_
FAIL After the mixer commanded
a change to the balance gas, the
status of the selector valve shows
the old balance gas is still
connected.
Mixer error bit
STS_CH2_DELTAP_FLOW_FAIL
Pressure difference between P3
and P2 differs from the drop
expected at the measured flow
for channel 2 (Balance Gas)
Screen
Mixer error bit
2
STS_CH1_DELTAP_FLOW_FAIL
Pressure difference differs from
the drop expected at the
measured flow for Channel 1 (O
Screen
Five seconds pass without
2
measured flow data from the
mixer.
Priority Source Enabling Criteria
Medium MGAS MGAS present and
MGAS communicates
this the continuous
occlusion alarm for 20
seconds
Medium MGAS MGAS present and
MGAS communicates
this the
OpenGasCircuit alarm
for 40 seconds
Medium MGAS
Medium
AC,
Mixer
Medium
AC,
Mixer
Medium
AC,
Mixer
)
2
Medium
AC
Communication has
been established
between mixer and AC.
11/03 1009-0357-000