Datex-Ohmeda S/5 Avance Technical Reference Manual page 146

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance
7-18
Test 7
Testing the bellows, the bellows pop-off valve, the bellows base
manifold, and the Bag/Vent switch
1. Separate the Bellows Module from the Circuit Module.
2. Insert appropriate test plugs into the bellows base manifold as shown to
the left.
Note: Position the bellows assembly so that the bellows remain collapsed
as you plug the ports.
Plug
3. Set Bag/Vent switch to Vent.
Plug
4. Position the bellows upright with the bellows collapsed.
5. Connect the Machine Test Tool to the interface ports as shown above.
6. On the Main Menu of the Service Application, select Gas Diagnostics.
7. On the Gas Diagnostics page, select O2 Flow.
8. Set O
Flow to 0.2 l/min.
2
9. On the Gas Diagnostics page, reselect the O2 Flow screen.
• Ensure that the Airway Pressure rises to ≥ 30 cm H
10.Set O2 Flow to OFF.
O.
2
11/03 1009-0357-000