Fresh Gas And O 2 Flush Flow (With Acgo) - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

2.11.5 Fresh gas and O
To ABS (Circle)
breathing system
(Figure 2-32)
ABS
ACGO
Selector
Switch
Figure 2-32 • Fresh gas and O
1009-0357-000 11/03
flush flow (with ACGO)
2
Fresh gas (1) flows from the vaporizer manifold outlet to the ACGO Selector Switch.
With the ACGO Selector Switch in the ABS position, fresh gas flow is channeled to the
breathing system.
The output of the O
Flush regulator (2) is channeled to the O
2
When activate, O
flush flow joins the fresh gas flow in the ACGO Selector Switch.
2
1
flush flow (to ABS)
2
2 Theory of Operation
Flush valve.
2
2
2-33