Inspiratory Flow Valve Cal; Bleed Resistor Cal - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

5.4.4 Inspiratory Flow Valve Cal

A
B
A

5.4.5 Bleed Resistor Cal

A
B
1009-0357-000 11/03
Calibration setup
Leave the Drive Port (A) and the Manifold Port (B) on the interface valve
plugged.
Calibration procedure:
1. On the Calibration menu, select Insp Flow Valve.
2. Push the ComWheel to enable the Stage 1 calibration.
3. When Stage 1 is completed, remove the plug from the Manifold port and
insert the calibrated orifice (C)
4. Push the ComWheel to enable the Stage 2 calibration.
(May take two minutes before you see any effects of the test on the screen.)
5. When Stage 2 is completed, select Previous Menu.
C
Calibration setup
1. Remove the Calibration Orifice from the Manifold port.
2. Plug the Manifold (B) port.
3. Leave the Drive Gas (A) port plugged.
Calibration procedure
1. On the Calibration menu, select Bleed Resistor.
2. Select Start.
3. When the test is completed, select Previous Menu.
C
Calibrated Flow Orifice
1504-3016-000
5 Calibration
5-13