Error Log Entry
Alarm Text
Action/Troubleshooting
INSP FLOW SENSOR
Replace insp flow
EEPROM FAILURE
sensor
Replace the Inspiratory Flow Sensor.
LOSS OF GAS
DELIVERY USER
SETTINGS
Reboot system. If problem continues, check the ACB to Mixer communication LED's (or VIB
communiction LED's).
1. If the RCV and XMT (or TXD and RXD) LED's indicate activity,
• check DU cable connections.
• replace Display Controller board if problem continues.
2. If the RCV and XMT ( or TXD and RXD) LED's indicate no activity,
• check the Anesthesia Contro board connection.
• replace the Anesthesia Control board if problem continues.
LOSS OF VENT
Vent Fail. Monitoring
PARAMETER SETTINGS
Only
Reboot system. If problem continues, check the ACB to Mixer communication LED's (or VIB
communiction LED's).
1. If the RCV and XMT (or TXD and RXD) LED's indicate activity,
• check DU cable connections.
• replace Display Controller board if problem continues.
2. If the RCV and XMT ( or TXD and RXD) LED's indicate no activity,
• check the Anesthesia Contro board connection.
replace the Anesthesia Control board if problem continues.
7-34
Condition (Basic info)
EEPROM cal data read failure
After regular communications
has been established between
the AC and the Display Computer,
this alarm shall be declared if the
system is in the Therapy State
and the AC determines the Gas
Delivery User Settings from the
Display Computer arrived more
than 10 seconds ago.
After regular communications
has been established between
the AC and the Display Computer,
this alarm shall be declared if the
system is in the Therapy State
and the AC determines the
Ventilator Parameter Settings
from the Display Computer
arrived more than 10 seconds
ago.
Priority Source Enabling Criteria
Low
AC,
Vent
AC
Medium
AC
11/03 1009-0357-000