Mixer Tests And Pressure - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance
8.2.3 Mixer Tests and
Pressure
Gas Diagnostics
Gas Supplies
Mixer Output
Mixer Tests and Pres
Mixer Temperatures
O2 Flow: 0.5
N2O Flow: OFF
Air Flow: OFF
Breathing Sys Leak
-> Main Menu
8-14
Selecting Mixer Tests and Pres brings up the Mixer Tests and Pres Gas
Delivery Diagnostics page.
M i x e r T e st s and P r e s G as Di ag n o sti cs
Last Power-Up Tests
O2 Proportional Valve Leak
Alt O2 Valve Leak
Balance Gas Prop Valve Leak
Balance Gas CheckValve Leak
O2 Flow Test
Balance Flow Test
Balance Gas ID
Pressure Data
O2 Pressure (P1)
O2 Pres Cal (P1)
Balance Pressure (P2)
Balance Pres Cal (P2)
Mixer Output Pres (P3)
Mixer Output Pres Cal (P3)
ADC Ref Voltage
Not done. No supply pressure
Not done. Selector valve incorrect state
Pass
Fail. Selector valve leaks
Fail. Proportional valve leaks
Pass; Fail
Not done. No supply pressure
Not done. Selector valve incorrect state
Pass
Fail. Selector valve leaks
Fail. Proportional valve leaks
Pass; Fail
Not done. No supply pressure;
Not done. Selector valve incorrect state;
Pass;
Fail, 3 l/min test;
Fail, 10 l/min test
Not done. No supply pressure;
Not done. Selector valve incorrect state;
Pass;
Fail, 3 l/min test;
Fail, 10 l/min test
None; Air; N2O
Value
Units
XX.XX
psi
X.XXX
Vdc
XX.XX
psi
X.XXX
Vdc
XX.XX
psi
X.XXX
Vdc
X.XXX
Vdc
11/03 1009-0357-000
kPa
XXX.XX
XXX.XX
XXX.XX