1009-0357-000 11/03
Test 8
Testing the bellows assembly
Note
If required, set up the Machine Test Tool and breathing system as shown in
Test 7.
1. Remove the bellows base manifold from the Bellows Module.
2. Insert appropriate test plugs into the bellows base manifold as shown to
the left.
Note: Position the bellows assembly so that the bellows remain collapsed
as you plug the ports.
Plug
3. Connect the tapered plug of the Machine Test Tool to the bellows base inlet
Plug
as shown to the left.
4. Position the bellows upright with the bellows collapsed.
Fresh Gas
5. On the Main Menu of the Service Application, select Gas Diagnostics.
6. On the Gas Diagnostics page, select O2 Flow.
7. Set O
Flow to 0.2 l/min.
2
8. On the Gas Diagnostics page, reselect the O2 Flow screen.
• Ensure that the Airway Pressure rises to ≥ 30 cm H
9. Set O2 Flow to OFF.
7 Troubleshooting
O.
2
7-19