Inspiratory Flow Valve - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance
4.4.6 Inspiratory Flow
Instructions
4-16
The Inspiratory Flow Valve instructions appear when the focus is on the Insp
Flow Valve menu item.
Valve
Read all steps before you start:
1. Complete the Manifold P Span calibration.
2. Put #2 plugs in the manifold and the drive gas ports of the vent engine.
3. Push the ComWheel to show the next menu.
4. Select Stage 1 calibration.
5. After Pass, replace the manifold port plug with the calibration orifice.
6. Select Stage 2 calibration.
7. You MUST do both stages for the calibration to be saved.
During calibration, a separate menu shows the counts and corresponding
flow at each step.
• If the outcome of both stages of the calibration is Pass, the new calibration
data is saved.
• If the outcome of either stage is Fail, the old calibration data is retained.
• The results of each stage of the calibration are saved to the Event Log.
Selecting Previous Menu before the calibration is done, aborts the calibration
in progress and keep the old calibration constants.