4.2 Install/Service Menu (Super User)

Show Alarm Limits
Save Default Case
Factory Defaults
1009-0357-000 11/03
Use the super-user password to access the Install/Service menu:
"16-4-34."
Menu Item
Message text
Colors
Set colors of parameters.
Units
Set units of weight, CO2, gas supply
pressure, and Paw.
Select yes to show alarm limits in
digit fields.
Save normal screen, air/ N2O,
circuit type, ventilator settings, and
alarm limits from the last case as
defaults
Return to default factory settings.
After selecting Factory Defaults:
"Reset machine for defaults to take
effect."
Installation
Set language, gas colors, hardware,
and enable options.
Service
Show technical data for
troubleshooting and calibration.
Exit
Turn power off to exit the service and
super user menus.
Install/Service
Trends Setup
Colors
Units
Show Alarm Limits
Save Default Case
Factory Defaults
Installation
Service
Exit
4 Installation and Service Menus
Comments
Default is Yes
Last used alarm settings (including hide/
show alarm limits), screen layout (middle
waveform selection, sweep speed,
scaling), ventilator mode and setting,
balance gas, and start case gas outlet
selection are saved as facility defaults.
Note: The Pmax alarm limit shall not be
saved higher than 40 cmH2O. The low
FiO2 alarm limit shall not be saved lower
than 21%.
Action: All facility defaults get replaced
with factory defaults. Super User settings
also get set to Factory Defaults. No
Service level configuration settings are
changed.
Navigate with password to Installation
menu. Password is "26-23-8"
Navigate with password to Service menu.
Password is "26-23-8"
Colors Menu
Units Menu
Paw
Weight
Flow
CO2
Gas Supply Pressure
Resp
Paw
CO2
Previous Menu
Previous Menu
4-3