Pipeline And Cylinder Tests; Supply Alarm Test - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance

3.5 Pipeline and cylinder tests

w w w w WARNING
3.5.1 O

supply alarm test

2
3-6
1. Connect the pipeline supplies one at a time and ensure that the
corresponding display indicates pipeline pressure.
2. Disconnect all pipeline supplies.
a. Open each cylinder valve.
b. Make sure that each cylinder has sufficient pressure. If not, close the
applicable cylinder valve and install a full cylinder.
3. Test the cylinder supplies for a high pressure leak. Make sure that each
cylinder has sufficient pressure:
a. If equipped, turn the auxiliary O
b. If equipped, turn off venturi derived suction.
c. Open each cylinder.
d. Record the cylinder pressure.
e. Close each cylinder valve.
f. Record the cylinder pressure after one minute. If the pressure
decreases more than indicated below, there is a leak.
5000 kPa (725 psig) for ventilator drive gas.
690 kPa (100 psig) for non ventilator drive gas.
If a cylinder supply fails this test, install a new cylinder gasket and do
this step again.
4. Close all cylinder valves.
Do not leave gas cylinder valves open if the pipeline supply is in use.
Cylinder supplies could be depleted, leaving an insufficient reserve
supply in case of pipeline failure
1. Establish O
, Air, and (if equipped) N
2
2. Set O
to 25% and (if equipped) N
2
N
O, set Air as balance gas.
2
3. Set total flow to 3 L/min.
4. Stop the O
supply. (Disconnect the pipeline supply or close the cylinder
2
valve.)
5. Make sure that:
a. The low "O
supply pressure low" alarm occurs.
2
b. The N
O (if equipped) and O
2
c. Air (if selected) flow continues or an Air selection prompt appears.
flow control fully clockwise (no flow).
2
.
0 gas supplies.
2
0 as balance gas. For machines without
2
flows stop.
2
11/03 1009-0357-000

Advertisement

Table of Contents
loading

Table of Contents