8.3.2 Vent Flow and
Pressure
Vent Diagnostics
Status
Vent Flow and Pres
Gas Inlet Vlv ON/OFF
Flow Valve Control
Toggle Circuit
-> Main Menu
8-20
Selecting Vent Flow and Pres brings up the Ventilation Flow and Pressure
page.
V e nt Flo w a nd P r e ssur e Di ag n o sti cs
Menu Item
Value
Gas Inlet Valve
Open or Closed
Flow Valve
XXXXX
Turn ComWheel to adjust flow valve
Circuit Command
Circle or Non-Circle
Inspiratory Flow
XXX.X
Expiratory Flow
XXX.X
Airway Pressure
XX X
Manifold Pressure
XXX
O2 Cell
XXX
ADC Ref Voltage
X.XXX
Flow Valve Setting
XXX.X
Flow Valve Feedback
xxxx
Flow Valve Current
xxxx
Units
Value
Counts
(appears when Flow Valve Control set ON)
l/min
XXXX
l/min
XXXX
cmH2O
XXXX
cmH2O
XXXX
%
XXXX
Vdc
l/min
XXXX
mV
XXXX
mA
XXXX