Copy Configuration - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Options List menu

4.3.4 Copy Configuration

Copy Configuration menu
Menu Item
Save to Card
Copy from Card
1009-0357-000 11/03
The options list shows which options are enabled.
Menu Item
Message text
Available Options
SIMV/PSV
SIMV vent w/pressure support.
PCV
Pressure controlled ventilation.
PSV Pro
Pressure support ventilation w/backup.
Message text
Save Configuration and
defaults to card.
Copy Configuration and
defaults from card.
When completed:
Copy from card complete.
Please reboot system.
Systems cannot accept configuration files from a different product model.
The software version is stored with the saved configuration. A system will
reject any configurations from other than the current version of software.
Selecting Save to Card overwrites any configuration on the card.
4 Installation and Service Menus
Values
<blank>, Fail, or OK.
The field is blank until the
data has either been
written to the card (OK) or
the system determines it
cannot write to the card
(Fail).
<blank>, Fail, or OK.
The field is blank until the
data has either been read
from the card (OK) or the
system determines it
cannot read the card or
the card does not have the
required data (Fail).
Values
On, Off
On, Off
On, Off
Comments
Saves all settings
that are not hardware
dependent, including
facility defaults,
colors, units, O2 flow
tube position,
decimal marker, and
altitude.
4-9

Advertisement

Table of Contents
loading

Table of Contents