Service Menu - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance

4.4 Service Menu

4-10
Use the service-level password to access the Service menu:
"26-23-8."
Whenever service menu is entered, "Enter Service dd-mmm-yyyy hh:mm:ss"
is recorded in the Event log.
Menu Item
Message text
SW HW versions
Scroll through system information.
Service Log
Show error, event, and alarm histories.
Calibration
Push ComWheel to perform service calibrations.
Previous Menu
Return to previous menu.
SW HW Versions
Total Time:
Software Release:
Model Code:
Machine Serial Number:
Option Package:
Option Code:
Anes Hardware . . .
Disp Hardware . . .
Mixer Hardware . . .
Vent Hardware . . .
Power Hardware . . .
MGas Hardware . . .
Service Log Menu
Calibration
Scroll Recent
Instructions
User Calibration
Error History
Manifold P Span
Event History
Insp Flow Zero
Alarm History
Insp Flow Valve
Copy Logs
Bleed Resistor
Reset Logs
Paw Span
Previous Menu
Zero Gas Xducer
Cal Config
Previous Menu
11/03 1009-0357-000