Configuring Case Defaults - Datex-Ohmeda Avance User's Reference Manual

Hide thumbs Also See for Avance:
Table of Contents

Advertisement

Avance
Configuring case
defaults
12-12
Set the user selectable default case types that show in the user's
Start Case menu through the Case Defaults menu. The Set to Last
Case selection is only available on submenus if a case has occurred.
Case Defaults
Edit/View Default:
Back
Available Defaults:
ADULT
PEDIATRIC
Patient and Sensor Type Adult
LOCAL
CUSTOM 1
Previous Menu
To change the settings for the default case types:
1. From Install/Service - Page 2, select Case Defaults.
Edit/View Default: is selected.
An arrow to the right of the default case name indicates the
selected case name.
2. Select the case name to enter the adjustment window.
3. To change the name of the case, select Name.
Select Clear to remove the existing name.
Select up to 10 characters from the list.
Select Delete to delete a character.
Select Save to save the name and close the selection
window.
Select Reset to return the name to the factory default name.
When the 10 character maximum is reached, the name is
automatically saved. The selection window closes.
4. Select another item to change. Make the change.
5. When finished setting the defaults, select Confirm.
6. Repeat to set the defaults for the other default case types.
7. Select Previous Menu to return to Install/Service - Page 2.
ADULT
Set to Last Case
Name ADULT
Patient Weight 70 kg
Patient Age 40 y
Other Gas Air
Circuit Circle
Data Source Vent
Vent Mode VCV
-More-
M1145956