3.1 Inspect the system

w w w w CAUTION
w w w w WARNING
3-2
The upper shelf weight limit is 34 kg (75 lb).
Do not leave gas cylinder valves open if the pipeline supply is in use.
Cylinder supplies could be depleted, leaving an insufficient reserve
supply in case of pipeline failure
Before testing the system, ensure that:
• The equipment is not damaged.
• Components are correctly attached.
• The breathing circuit is correctly connected, not damaged.
• Pipeline gas supplies are connected.
• Cylinder valves are closed.
• Models with cylinder supplies have a cylinder wrench attached to the
system.
• Models with cylinder supplies have a reserve supply of O
the machine during system checkout.
• The casters are not loose and the brakes are set and prevent movement.
• The power cord is connected to a wall outlet. The mains indicator comes
on when AC Power is connected.
.
2
11/03 1009-0357-000
connected to