Datex-Ohmeda S/5 Avance Technical Reference Manual page 95

Anesthesia machine
Table of Contents

Advertisement

Insp Flow Valve Menu
Insp Flow Valve Data menu
1009-0357-000 11/03
Menu Item
Message text
Stage 1
Calibrate the insp flow valve at low
flows.
Blocked text:
"Connect a supply of the drive gas to
continue."
Stage 2
Calibrate the insp flow valve at high
flows.
Blocked text:
"Stage 1 calibration is required
first."
Insp Flow Valve Data
Show insp flow valve calibration
table.
Previous Menu
Return to previous menu.
During Calibration:
"Calibration in progress. Push
ComWheel to cancel."
The Insp Flow Valve Data menu contains a table of 24 entries from the
previous calibration. The table is erased at the start of Stage 1. The table is
update in real time during the calibration.
4 Installation and Service Menus
Comments
Blocked if the
ventilator drive
gas supply
pressure would
cause a gas
supply alarm
during normal
operation.
Blocked if Stage
1 has not been
completed.
4-17