Manual Ventilation - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

2.11.2 Manual ventilation

Manual inspiration
(Figure 2-24)
AP Airway Pressure
B Bag/Vent switch to Bag
FG Fresh Gas
1 Flow to absorber
2 Flow from absorber
3 Inspiratory flow
Figure 2-24 • Gas flow during manual inspiration
1009-0357-000 11/03
The Bag/Vent switch closes the ventilator path (B). . . .
Gas flows from the bag (1), through the absorber (2), into the breathing circuit module,
and through a unidirectional valve (inspiratory check valve) to the patient (3).
During inspiration, fresh gas (FG) flows from the machine into the inspiratory limb,
upstream of the inspiratory check valve.
FG
B
1
2 Theory of Operation
AP
3
2
1
3
2-25