Replace Vaporizer Pressure Relief Valve - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance
9.12.4 Replace
vaporizer pressure
relief valve
9-36
1. Set the system switch to Standby.
2. Remove the vaporizers from the vaporizer manifold.
3. Remove the upper rear panel (Section 9.2).
4. Using a 13mm open ended wrench, remove the vaporizer pressure relief valve by
turning counterclockwise.
5. Verify that the parts are free of dust and dirt.
6. Install a new vaporizer pressure relief valve.
7. To reassemble, perform the previous steps in reverse order.
8. Perform the checkout procedure (Section 3).
11/03 1009-0357-000