Units Menu; Options Key - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance

4.3.2 Units Menu

4.3.3 Options Key

Options Key menu
4-8
This is the same menu that is accessible from the super-user Install/Service
menu.
Menu Item
Message text
Weight
Change weight unit: kg or lb.
CO2
Change CO2 unit: %, kPa, or mmHg.
Gas Supply Pressure
Change gas supply pressure unit:
kPa, psi, or bar.
Paw
Change Paw unit: kPa, hPa, cmH2O,
mmHg, mbar.
The Options Key menu is used to configure the software to include the
features that the customer has purchased. The included features are shown in
the Options List menu.
Menu Item
Message text
Current Key
Enter key code to enable options.
Entry 1
Enter first entry of key-code.
Entry 2
Enter second entry of key-code.
Entry 3
Enter third entry of key-code.
Entry 4
Enter fourth entry of key-code.
Entry 5
Enter fifth entry of key-code.
Entry 6
Enter sixth entry of key-code.
Entry 7
Enter seventh entry of key-code.
Save New Key
Confirm entries for key-code.
Control Board ID
Control number used by key-code.
When options are added, "Add <option> dd-MMM-yyy hh:mm:ss" is written to
the event log.
If more than one option is added, each option is be listed separately.
Values
kg or lb
%, kPa,
or mmHg
psi, kPa, bar
kPa, hPa,
cmH2O, mmHg,
or mbar
Values
XXXAXBC
0 to 9, A to Z, ~,
!, @, #, $, %, ^,
*, (,), ?
XXX
11/03 1009-0357-000