Error Log Entry
Alarm Text
Action/Troubleshooting
BATTERY EMPTY
System shutdown in
<5 min
Leave the system plugged in to charge the battery.
If problem continues, check the battery charge circuit in Service Software.
Replace Battery.
BATTERY FAIL
No Battery Backup.
Leave the system plugged in to charge the battery.
If problem continues, check the battery charge circuit in Service Software.
Replace Battery.
BATTERY LOW
Plug in power cable.
On battery
Leave the system plugged in to charge the battery.
If problem continues, check the battery charge circuit in Service Software.
Replace Battery.
BATTERY MISSING
No battery backup
Connect Battery
Leave the system plugged in to charge the battery.
If problem continues, check the battery charge circuit in Service Software.
Replace Battery.
BATTERY REVERSED
No battery backup
CONNECTIONS
Check Battery Connections
BATTERY V LOW
Plug in power cable.
On battery
Leave the system plugged in to charge the battery.
If problem continues, check the battery charge circuit in Service Software.
Replace Battery.
1009-0357-000 11/03
Condition (Basic info)
Available battery power is
between 1 and 5 minutes
Battery voltage <10.5 V
or
While in bulk, over, or float
charging battery is <10.5VDC
or
Battery has been bulk charging
for >12 h in Standby or 24 h while
powered on.
Or
Voltage > 15.5V during bulk or
over charging and normal current
>0.25 Amps
Available battery power
decreases to between 20 and 30
min
Any battery voltage is between
±1.0 VDC.
Any battery voltage is less than –
1.0 VDC
Available battery power
decreases to between 10 and 20
minutes
7 Troubleshooting
Priority Source Enabling Criteria
High
PC
AC Mains Power Failure
in progress.
Medium
PC
Medium
PC
Mains AC Mains Power
Failure in progress.
Medium
PC
POST state
Medium
PC
Medium
PC
AC Mains Power Failure
in progress.
7-31