Datex-Ohmeda S/5 Avance Technical Reference Manual page 170

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance
Error Log Entry
Alarm Text
Action/Troubleshooting
REVERSE EXPIRATORY
Reverse exp flow.
FLOW
Check valves OK?
Check flow sensor connections for "No Flow Sensor" alarm.
Check the breathing circuit.
Perform flow sensor calibration.
Check Insp/Exp check valves.
Replace the flow sensors.
Check for kinked VIB tubing.
Check the VIB cabling.
SCGO
Vol and Apnea
monitoring off
No service action required.
STANDBY PATIENT
No fresh gas flow!
DETECTION
No service action required.
VENT FLOW VALVE FAIL
Vent Fail. Monitoring
DAC
Only
Reboot System. If problem continues, in the Service Software / Vent Flow & Pressure Diagnosis,
increase the Flow Valve counts and view the Flow Valve Feedback mV and Counts. Verify the settings
match.
VENT +12.5V FAIL
Vent Fail. Monitoring
Only
In the Service Software / Vent Interface Bd Power Diagnosis, view the "Vent Int Bd 10VA Voltage" from
Board Supplies:
If "Vent Int Bd 10VA Voltage" reads "OK, and +12.5 Vdc reads "Fail", replace the VIB.
If "Vent Int Bd 10VA Voltage" reads "Fail" and the "Vent Int Bd 10VA Voltage" from the Anes Cntrl Bd
reads "OK", Check cabling between ACB and VIB.
7-42
Condition (Basic info)
Flow towards the patient (volume
>= 20 mL) on expiratory sensor
and flow towards the patient
(volume >= 5 mL) on the
inspiratory sensor during
inspiration for 6 consecutive
mechanical breaths
Non Circle SCGO selected
3 volume breaths are detected
within 30 seconds or 3 CO
2
breaths are detected within 30
seconds
Incorrect DAC feedback for 3
consecutive readings
Nominal 12.5V <11.3 Vdc or
>13.13 Vdc
Priority Source Enabling Criteria
Medium AC Vent In-range flow data
available, mechanical
ventilation on
Low
DC
System has SCGO
High
DC
System in Checkout:
General or Checkout:
Start Case
Medium
AC,
Vent
Medium
AC,
Vent
11/03 1009-0357-000