Datex-Ohmeda S/5 Avance Technical Reference Manual page 171

Anesthesia machine
Table of Contents

Advertisement

Error Log Entry
Alarm Text
Action/Troubleshooting
VENT +6V FAIL
Vent Fail. Monitoring
Only
In the Service Software / Vent Interface Bd Power Diagnosis, view the "Vent Int Bd 10VA Voltage" from
Board Supplies:
If "Vent Int Bd 10VA Voltage" reads "OK, and +6.0Vdc reads "Fail", replace the VIB.
If "Vent Int Bd 10VA Voltage" reads "Fail" and the "Vent Int Bd 10VA Voltage" from the Anes Cntrl Bd
reads "OK", Check cabling between ACB and VIB.
VENT 1.22V FAIL
Vent Fail. Monitoring
Only
In the Service Software / Vent Interface Bd Power Diagnosis, view the "Vent Int Bd 10VA Voltage" from
Board Supplies:
If "Vent Int Bd 10VA Voltage" reads "OK, and 1.22 Vdc reads "Fail", replace the VIB.
If "Vent Int Bd 10VA Voltage" reads "Fail" and the "Vent Int Bd 10VA Voltage" from the Anes Cntrl Bd
reads "OK", Check cabling between ACB and VIB
VENT -6V FAIL
Vent Fail. Monitoring
Only
In the Service Software / Vent Interface Bd Power Diagnosis, view the "Vent Int Bd 10VA Voltage" from
Board Supplies:
If "Vent Int Bd 10VA Voltage" reads "OK, and
If "Vent Int Bd 10VA Voltage" reads "Fail" and the "Vent Int Bd 10VA Voltage" from the Anes Cntrl Bd
reads "OK", Check cabling between ACB and VIB.
VENT ADC VREF FAIL
Vent Fail. Monitoring
Only
In the Service Software / Vent Interface Bd Power Diagnosis, view the "Vent Int Bd 10VA Voltage" from
Board Supplies:
If "Vent Int Bd 10VA Voltage" reads "OK, and 3.2 Vdc reads "Fail", replace the VIB.
If "Vent Int Bd 10VA Voltage" reads "Fail" and the "Vent Int Bd 10VA Voltage" from the Anes Cntrl Bd
reads "OK", Check cabling between ACB and VIB.
VENT AIRWAY
Inspiration stopped
OVERPRESS SIGNAL
No Service Action.
Reboot system. If problem continues, check Airway Pressure signal in Service Mode.
VENT AIRWAY
Vent Fail. Monitoring
OVERPRESS SIGNAL
Only
FAIL
No Service Action.
Reboot system. If problem continues, check Airway Pressure signal in Service Mode.
VENT FLOW VALVE FAIL
Vent Fail. Monitoring
CURRENT
Only
Reboot System. If problem continues, in the Service Software / Vent Flow & Pressure Diagnosis,
increase the Flow Valve counts and view the Flow Valve Current mA and Counts.
VENT SUSTAINED PAW
Vent Fail. Monitoring
SDOWN
Only
No Service Action.
Reboot system. If problem continues, check Airway Pressure signal in Service Mode.
1009-0357-000 11/03
Condition (Basic info)
VSIB +6V out of range<5.51 Vdc
or > 6.5 Vdc
Voltage < 1.074Vdc or Voltage >
1.367 Vdc
VSIB -6V out of range<-6.72 Vdc
or > -5.28 Vdc
-6.0Vdc reads "Fail", replace the VIB.
VSIB ADC3.200V ref voltage out
of range <3.179 or >3.221 Vdc
High airway overpressure signal
set.
Ventilator SIB indicates the High
Airway overpressure signal was
set and Paw < 90 cmH
O and
2
Pmanifold <80 cm H
O.
2
Incorrect current feedback for 7
consecutive readings
Paw > 100 cmH
O for 10
2
seconds
7 Troubleshooting
Priority Source Enabling Criteria
High
AC,
Vent +12.5 V (10 VA) is
Vent
OK
Medium
AC,
Vent +12.5 V (10 VA) is
Vent
OK
High
AC,
Vent +12.5 V (10 VA) is
Vent
OK
High
AC,
Vent +12.5 V (10 VA) is
Vent
OK
Medium
AC,
Mechanical Ventilation
Vent
On
Medium
AC,
Mechanical Ventilation
Vent
On
Medium
AC,
Vent
Medium
AC,
In-range Paw data
Vent
available
7-43