Datex-Ohmeda S/5 Avance Technical Reference Manual page 138

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance
Leak in
Flow Sensor Module
or Circuit Module
Perform Tests 14 and 15
Testing the Flow Sensors
Fail
Pass
using the low-pressure
leak test device
Check/Replace
Check/Replace the seals
Flow Sensors or
on the Circuit Module and
bulkhead connector
the O-Ring on the O
cell
2
O-Rings
Swap Module(s)
Do System
with another machine.
"All Checks"
Fail
Pass
Done
Repair or replace Module
(Section 3.2)
5
Flowchart
7-10
11/03 1009-0357-000

Advertisement

Table of Contents
loading