9.9.3 Replace the
inlet manifold
9-28
1. Remove the rear panel (Section 9.2).
2. Disconnect the tubing from the manifold outlet(s).
3. Remove the two screw that hold the manifold to the side extrution.
3
2
4
4. Transfer the following item to the replacement manifold or install new as required.
• pipeline check valve (1)
• inlet filter (2)
• inlet fitting (3) and o-ring (4)
• relief valve (5)
5. Transfer the pressure transducer to the new supply (Section 9.11).
• Remove any teflon tape remnants from the transducer mounting threads
(transducer and module).
• Apply 1-1/4 turns of new teflon tape around the treads. Verify that the first few
threads are free of tape.
• Install the transducer.
6. To reassemble, perform the previous steps in reverse order.
7. Perform the checkout procedure (Section 3).
1
5
11/03 1009-0357-000