Anesthesia Control Board Power Diagnostics - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance
8.1.4 Anesthesia Control
Board Power Diagnostics
Power Diagnostics
Power Control
Anes Control Board
Mixer
Vent Interface Bd
Display Unit
->Main Menu
8-6
There are two pages of diagnostics for the Anesthesia Control board.
Selecting Anes Control Board brings up the first page of the Anesthesia
Control Board Power Diagnostics.
An e s C o n tr o l Bo ar d P o w e r Di ag no st i cs
Label
12.5Vdc from Pwr Cntrl
ADC Reference
Gas Select 10VA Volts
Gas Select 10VA Amps
P Xducer 10VA Amps
Vent Int Bd 10VA Volts
Vent Int Bd 10VA Amps
Vent Valves 10VA Volts
Vent Valves 10VA Amps
Acces 1 10VA Volts
Acces 1 10VA Amps
Turn the ComWheel to select the second page.
Push the ComWheel to return focus to the Power Diagnostics selection menu.
( Pa g e 1 o f 2 )
Value Format
Units
XX.XX
Vdc
X.XXX
Vdc
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
OK, Fail
Normal range
11.90 to 12.90
4.018 to 4.176
11/03 1009-0357-000