Datex-Ohmeda S/5 Avance Technical Reference Manual page 60

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance
Auxiliary (Non-circle)
Common Gas Outlet
(Figure 2-33)
O
Cell
2
O
Sense
2
ACGO
ACGO
Selector
Switch
Figure 2-33 • Fresh gas and O
2-34
Fresh gas (1) flows from the vaporizer manifold outlet to the ACGO Selector Switch.
With the ACGO Selector Switch in the ACGO position, fresh gas flow is channeled to the
ACGO outlet.
At the ACGO outlet, a small sample is diverted to the O
monitoring.
The output of the O
Flush regulator (2) is channeled to the O
2
When activated, O
flush flow joins the fresh gas flow in the ACGO Selector Switch.
2
1
flush flow (to ACGO)
2
Cell in the ABS for O
2
2
Flush valve.
2
2
11/03 1009-0357-000