Table of contents TABLE OF CONTENTS Command Boards / Bars, Remote Controller TABLE OF CONTENTS Introduction Specifications 1.1 Command Board, K-ANE/K-ICU ........................2 1.2 Command Bar, K-ANEB/K-ICUB ........................2 1.3 Remote Controller, K-REMCO ........................2 Functional Description 2.1 Command Boards, K-ANE/K-ICU and Command Bars, K-ANEB/ K-ICUB .............3 2.1.1 Connectors and signals ........................4 2.2 Remote Controller, K-REMCO ........................5 Service procedures...
Page 4
S/5 Keyboards 6.4.4 Front panel stickers for K-ANEB/ICUB .....................21 Earlier Revisions APPENDICES A, B Service check form Command Board, K-ANE/K-ICU, Command Bar, K-ANEB/K-ICUB ........A-1 Service check form Remote Controller, K-REMCO ...................B-1 Document No. 8001589-1...
Command Boards / Bars, Remote Controller INTRODUCTION This section provides information about the maintenance and service of the following products: Command Boards, K-ANE/ K-ICU, • Command Bars, K-ANEB/ K-ICUB • Remote Controller, K-REMCO • Document No. 8001589 -1...
S/5 Keyboards SPECIFICATIONS 1.1 Command Board, K-ANE/K-ICU Dimensions, W × D × H 315 × 55 × 43 mm Weight (incl. cable) 0.5 kg Input voltage Power consumption 450 mW max Communication protocol RS-232 NOTE: Power supply from the display controller board only. 1.2 Command Bar, K-ANEB/K-ICUB Dimensions, W ×...
S/5 Keyboards LEDs The alarm LEDs are activated by the Command Board/Bar PCB under the commands received via serial communication from the CPU board. The red or yellow alarm LED is lit when red or yellow alarm is activated. STBY -led is lit when the device is turned to stand-by and connected to mains. NOTE: If there are two Command Boards/Bars connected to the system, the system is ON (STBY LED not lit), when at least one of the switches is positioned ON.
Command Boards / Bars, Remote Controller 2.2 Remote Controller, K-REMCO The Remote Controller consists of 12 direct function keys and the ComWheel. K-REMCO PCBs The K-REMCO has two PCBs located inside the Remote Controller. One board has only the pushbutton switches of the keys. The other board reads the status of the keys and the ComWheel and forwards the information to the CPU board.
Field service is limited to replacing faulty PC boards or mechanical parts. The PC boards are then returned to Datex-Ohmeda for repair. Datex-Ohmeda Technical Services is always available for service advice. Please provide the unit serial number, full type designation and a detailed description of the fault.
Page 11
Command Boards / Bars, Remote Controller Check that the plastic front cover and the front panel sticker are intact. Check that the ComWheel cover is intact and is attached properly. Check that the ON/STBY -switch and its protector (in K-ANE/K-ICU) are intact and are attached properly.
S/5 Keyboards 11. Check that the Command Board/Bar functions normally after the performed electrical safety check. 12. Clean the the Command Board/Bar. Fill in all necessary documents. • 3.2.3 Remote Controller, K-REMCO Turn the monitor to STBY. • Disconnect the remote controller cable from the display controller board. •...
Page 13
Command Boards / Bars, Remote Controller Wait until normal monitoring screen appears, then check that the picture on the screen is displayed with correct resolution. If the resolution is not correct, replace the remote controller cable. Enter the service menu: Menu - Monitor Setup (on the remote controller keypad)
S/5 Keyboards 3.3 Disassembly and reassembly 3.3.1 Command Board, K-ANE/ K-ICU The Command Board is disassembled according to the following procedure. Please refer to the exploded view of the Command Board. Disconnect the display power cable, the display data cable and the Command Board cable from the Central Unit.
Command Boards / Bars, Remote Controller 3.3.3 Remote Controller K-REMCO The Remote Controller is disassembled according to the following procedure. Please refer to the exploded view of the Remote Controller (chapter 6, spare parts). Disconnect the K-REMCO cable from the Central unit. Pull out the knob of the ComWheel.
S/5 Keyboards TROUBLESHOOTING 4.1 K-ANE/ K-ICU/ K-ANEB/ K-ICUB/ K-REMCO See Keyboard Service Menu in chapter 5, and perform tests available. If any of the tests fail, see explanation below. Problem Cause Treatment Keyboard cable loose or broken. D-26 ON/STBY switch not working Check the items.
S/5 Keyboards 5.1 Keyboard menu The service menu for testing the command board functions. Upper Led is for testing the upper alarm LED (red) on the command board. When the text is highlighted, the upper alarm LED can be turned on and off by pressing the ComWheel.
Command Boards / Bars, Remote Controller 5.1.2 Keyboard Type Store Mask A selection for setting the anaesthesia keyboard’s language. The selected language determines the outcome of the lower keypad. Store Type is for setting the keyboard’s type; COM = Command Board ARK = Anaesthesia Keyboard AIC = Information Center Keyboard NOTE: The settings should be checked if the...
S/5 Keyboards SPARE PARTS NOTE: Only changed part numbers are listed under later revisions. To find the desired part: check first the list of the revision that corresponds your device. If the part is not listed there, check the previous revision, etc. until you find the right number. * this part is recommended for stock Item numbers refer to the exploded view.
Page 26
S/5 Keyboards EARLIER REVISIONS For more service information on the earlier revisions, please refer to: Command Board revision 01 Service Manual p/n 880850 Command Board revision 02 Service Manual p/n 885930 This manual supports all later revisions. Document No. 8001589-1...
Page 27
Appendices A, B, Service check forms, K-ANE/ICU, K-ANEB/ICUB, K-REMCO APPENDICES A, B Document No. 8001589 -1...
Page 29
Appendices A, B, Service check forms, K-ANE/ICU, K-ANEB/ICUB, K-REMCO Service check form Command Board, K-ANE/K-ICU, Command Bar, K-ANEB/K-ICUB Customer Service Keyboard type Service engineer Date OK = Test OK N.A. = Test not applicable Fail = Test Failed N.A. Fail N.A.
Page 31
Appendices A, B, Service check forms, K-ANE/ICU, K-ANEB/ICUB, K-REMCO Service check form Remote Controller, K-REMCO Customer Service Keyboard type Service engineer Date OK = Test OK N.A. = Test not applicable Fail = Test Failed N.A. Fail N.A. Fail 1. Internal parts 2.