Page 1
Datex-Ohmeda Light Monitor, F-LM1, F-LMP1 Technical Reference Manual All specifications are subject to change without notice. CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner. Outside the USA, check local laws for any restriction that may apply.
Page 2
Intended purpose Datex-Ohmeda S/5 Light Monitor, F-LM1, F-LMP1 is intended for stationary and mobile monitoring of patient’s ECG, SpO2, body temperature, invasive and non-invasive blood pressures as well as respiration and CO2 in hospitals and ambulances. When used outdoors, the monitor must be in the protective case. The monitor is intended for use by qualified medical personnel only.
Page 3
Unit 8000060-1 Isolation Power Supply Unit for CO Unit Datex-Ohmeda Light Monitor Recorder, N-LREC 890201-5 Datex-Ohmeda Light Monitor Power Adapter, N-LPOW M1051340 Datex-Ohmeda Light Monitor Battery Module, F-LBAT 890200-5 Datex-Ohmeda Light External Battery Charger, N-LCHGR 896192-1 DataCard and Network Options, N-LDATA, N-LNET, N-LDNET...
Table of contents TABLE OF CONTENTS PART 1 - GENERAL SERVICE GUIDE TABLE OF CONTENTS Part 1 - General Service Guide TABLE OF FIGURES INTRODUCTION Notes to the reader..........................2 About this manual..........................2 Related documents ..........................2 Installation ............................3 1.4.1 Choosing the location ........................
Page 6
Datex-Ohmeda S/5 Light Monitor 5.1.4 ECG board tests......................... 25 5.1.5 STP board tests ......................... 25 5.1.6 NIBP board test ......................... 26 5.1.7 Mainstream CO test........................26 Performance checks........................... 27 SERVICE CHECK General service information........................ 30 Service check procedure ........................30 6.2.1...
Table of contents 10.7 External power adapter........................57 10.8 Light External Battery Charger (optional)..................58 10.9 Output specifications ........................58 10.9.1 Nurse Call and Defibrillator Synchronization Signals..............58 APPENDICES FUNCTIONAL CHECK form SERVICE CHECK form EMC Guidance TABLE OF FIGURES Figure 1 S/5 Light Monitor with battery module and power adapter ..............1 Figure 2...
The S/5 Light Monitor is a portable factory configurated compact monitor designed for stationary and mobile indoor monitoring of a patient’s vital signs. The monitor is available in two configurations; F-LM1 and F-LMP1. The difference between the configurations is illustrated in table 1.
Datex-Ohmeda S/5 Light Monitor The following factory configurable parameters and functions are available: • N-LNET; Network option • N-LDATA; DataCard option • N- LDNET; Network and DataCard options • N-LCM; Mainstream CO option To all monitors following options are available: •...
Introduction 1.4 Installation 1.4.1 Choosing the location The monitor can be placed on a flat surface or hung with the handle from the bed or wall rails. Make sure that the surface or rail holds up to at least 10 kg/22 lb. 1.4.2 Warnings WARNING The monitor or its components should not be used adjacent to or stacked with...
Datex-Ohmeda S/5 Light Monitor 1.5 Symbols 1.5.1 Symbols on equipment Attention, consult accompanying documents. on the rear panel Electric shock hazard. Do not open monitor frame. Refer servicing to qualified personnel. indicates the Do not touch any part of monitor or patient connections during defibrillation procedure.
Page 13
Introduction : Type BF (IEC-60601-1) protection against electrical shock, not defibrillation proof. : Type CF (IEC-60601-1) protection against electrical shock, not defibrillation proof. NIBP: Type BF (IEC-60601-1) defibrillator-proof protection against electrical shock. ECG, P : Type CF (IEC-60601-1) defibrillator-proof protection against electrical shock. Battery packs contain heavy metal cadmium (chemical symbol Cd) and, in case of disposal, must be separated from other waste according to local regulations.
Datex-Ohmeda S/5 Light Monitor IPX1 – vertically falling water drops shall have no harmful effects. IPX4 – splash proof SN, S/N Serial number Battery charging LEDs: LEDs will blink while batteries are being charged, and will stay illuminated when batteries are fully charged and the monitor is connected to mains.
Page 15
Introduction This symbol is to indicate the temperature limitations within which the transport package should be kept and handled. Document No. M1051339...
Datex-Ohmeda S/5 Light Monitor SAFETY 2.1 Safety precautions 2.1.1 Warnings WARNING Indicates situations in which the user or the patient may be in danger of injury or death. Explosion hazard • Do not use the monitor in presence of flammable anesthetics.
Page 17
Safety Patient safety • Always make sure that necessary alarm limits are set and operative when you start monitoring. • When the alarms are suppressed, observe the patient frequently. • Connect only one patient to one monitor at a time. •...
Page 18
Datex-Ohmeda S/5 Light Monitor WARNING Use only approved accessories, mounts and defibrillator-proof cables and invasive pressure transducers. For a list of approved supplies and accessories, see the Supplies and Accessories catalog. Other cables, transducers and accessories may cause a safety hazard, damage the equipment or system, result in increased emissions or decreased immunity of the equipment or system or interfere with the measurement.
Indicates situations in which the unit or devices connected to it may be damaged. • Use licensed patient cables and accessories only, see Datex-Ohmeda Catalogue Patient Monitor Supplies and Accessories. Other cables and accessories may damage the system or interfere with measurement.
Datex-Ohmeda S/5 Light Monitor Storage and transport • For allowed storage and transport conditions refer to the documentation delivered with the monitor. Points to note • Medical electrical equipment needs special precautions regarding electromagnetic compatibility and needs to be installed and put into service by qualified personnel according to the electromagnetic compatibility information provided in the Appendix C.
Safety 2.3 Safety test Measure leakage current according to following procedures: • Connect the GND electrode of the test device to the body of the 9-pin D-connector of the Light Monitor. The nuts that are for locking the male cable connector can be used as well. Document No.
Datex-Ohmeda S/5 Light Monitor ARCHITECTURE The Light Monitor is a compact patient monitor with fixed configuration. The monitor is designed for stationary and mobile indoor monitoring of a patient’s vital signs. The monitor can be equipped with an optional recorder and battery module. An external power adapter is provided for power supply.
Page 23
Architecture STP unit The STP unit is a parameter board that measures invasive blood pressure(s), body temperature and oxygen saturation of the blood (SpO2). ECG unit The ECG unit is a parameter board that measures 3-lead ECG. NIBP unit The NIBP unit is a parameter board that measures non-invasive blood pressure. unit The CO unit measures carbon dioxide consentration of the respiration gases.
Datex-Ohmeda S/5 Light Monitor 3.2 Power supply unit The power supply board functions can be summarized as follows: • Charge the batteries. • Choose between battery and mains operation. • Distribute power to the CPU, CO , STP, ECG, and NIBP boards.
Architecture 3.4 NIBP unit The NIBP unit consists of the NIBP board, an air pump, and pneumatic valves. The NIBP board functions can be summarized as follows: • Control the function of the air pump and the pneumatic valves. • Measure the pressure variation of the NIBP cuff.
3.10.1 Module bus The module bus is an asynchronous, CMOS-level serial communication between host processor and Datex-Ohmeda based parameter modules and CO measurement by Welch Allyn. Only TXDM and RXDM signals are used. The TXDM signal is module bus transmit line from host to modules and the RXDM signal is a receive line from modules to host.
Architecture 3.10.3 CO module The CO module, manufactured by Welch Allyn is used in the Light Monitor for CO mainstream measurement. Communication with the host is provided by an asynchronous serial link. Signals used in the communication are TXDC, RXDC. The TXDC signal is transmit line from host to CO module and the RXDC signal is a receive line from CO module to host.
Datex-Ohmeda S/5 Light Monitor CLEANING Before cleaning: • Turn the power to Standby • Disconnect the power cord from the mains power supply After cleaning: • Let dry completely • Reconnect the power cord and turn the power ON CLEANING CASING /...
Functional check FUNCTIONAL CHECK These instructions include complete procedures for functional check (FC) for Datex-Ohmeda Light Monitors. The functional check is recommended to be performed after monitor installation and after long storage. The first part of the instructions contains check procedures that are done through the service menu.
Page 30
Datex-Ohmeda S/5 Light Monitor Switch the monitor on. Check that the monitor starts up properly, i.e. − both alarm LED’s turn on and off − the fan starts running − the start-up sound is heard from the loudspeaker − the normal monitoring screen appears −...
Functional check Reconnect the power cord and check that during charging, the charging symbol is displayed and the Battery charge status LED starts flashing: 5.1.1 Recorder test Open the paper compartment cover. Check that the message “Recorder: Cover open” appears on the screen, then close the cover. Select Main Menu - Record/Print- Record Wave and check that the recorder starts recording.
5.1.3 Network test Connect the Mon-Net cable and the Identification plug to the monitor. Check that the monitor connects to the Datex-Ohmeda Network, i.e. the network symbol is displayed on the upper right-hand corner of the screen. A message regarding the connection to Datex-Ohmeda Information Center should appear in the message field on the screen.
Functional check 5.1.4 ECG board tests Enter the ESTP : ECG service menu: Menu Press the button and select: Monitor Setup - Install/Service (Password 16-4-34) - Service View (Password 26-23-8) - Modules - ESTP : ECG Check from the Service Data that: −...
Datex-Ohmeda S/5 Light Monitor 5.1.6 NIBP board test Enter the NIBP service menu: Check that the “Timeouts”, “Bad checksums” and “Bad c-s by mod” values are not increasing faster than by 50 per second. Check that the NIBP board memories have passed the internal memory test, i.e. “RAM”, “ROM”...
Functional check 5.2 Performance checks Connect a 5-lead cable to the monitor. Configure the monitor screen so that all parameter information is displayed, for example for the maximum configuration: Menu Press the button to enter the Main Menu and select: Monitor Setup - Waveform Fields/Digit Fields Waveform Fields Digit Fields...
Page 36
Datex-Ohmeda S/5 Light Monitor Connect the patient simulator to the ECG and InvBP connectors, SpO probe to the SpO connector and temperature test plug to the Temp connector. Attach the SpO probe on your finger and check that: − the simulated waveforms are good −...
Page 37
Functional check Check that the message “No CO2 sensor” is displayed when there is no CO sensor connected. Connect a CO sensor to the monitor and check that the message “warming” is displayed. Wait until the “warming” message disappears and the CO value appears on the screen.
Datex-Ohmeda S/5 Light Monitor SERVICE CHECK 6.1 General service information Field service of the Light Monitor is limited to replacing faulty printed circuit boards or mechanical parts. Faulty printed circuit boards should be returned to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the unit serial number, full type designation, and a detailed description of the fault.
Service check Tool Order No. CO2 option) Mainstream CO2 airway adapter (only for monitors 902301/902302/ with CO2 option) 902303 Screwdriver -Pozidrive- type MemCard- Data (only for monitors with data card option) Recommended parts Part Order No. Notes Fan filter 886841 Recorder paper 74205 if N-LREC option...
Page 40
Datex-Ohmeda S/5 Light Monitor Figure 2 Adjustment trimmer R23 in N-LPOW − Clean or replace the fan filter. − Clean the recorder unit, if necessary. − Check that all shielding flaps (6 pcs) are properly closed. (Figure 3) Shielding Flaps...
Page 41
Service check both screws on the patient panel are tightened properly − General functional inspection − Connect the power cord to the Power Adapter and the Power Adapter cable to the Light Monitor. Check that the green LED turns on in the power adapter and the LED on the lower right-hand corner of the monitor turns on or starts flashing.
Page 42
Datex-Ohmeda S/5 Light Monitor Preset ECG, Respiration, InvBP and SpO measurement settings: ECG & Resp ECG Setup - Hr Source - Auto - Pacemaker - Show - Resp Setup - Size - 1.0 - Resp Rate Source - Auto - Measurement - On - Detection Limit - Auto NIBP &...
6.2.4 Network test Connect the Mon-Net cable and the Identification plug to the monitor. Check that the monitor connects to the Datex-Ohmeda Network, i.e. the network symbol is displayed on the upper right-hand corner of the screen. A message regarding the connection to Datex-Ohmeda Information Center should appear in the message field on the screen.
Datex-Ohmeda S/5 Light Monitor 6.2.5 ECG measurement Enter the service menu: Menu - Monitor Setup - Install/Service (Password 16-4-34) - Service View (Password 26-23-8) Take down the information regarding monitor software and unit id by selecting Scroll Vers and turning the ComWheel or choose Record Vers if the recorder is installed.
Service check Lead selection Lead to be detached Message Lead I L/LA Leads off R/RA Leads off N/LL Leads off Lead II L/LA Leads off R/RA Leads off N/LL Leads off Lead III L/LA Leads off R/RA Leads off N/LL Leads off Check that in any case the correspondence electrode (RA, LA or LL) shows off.
Page 46
Datex-Ohmeda S/5 Light Monitor RATIO - 1/1 APNEA - OFF SHIFT – OFF Check that the Resp waveform is shown and the RR value is 20 (±5). Change the position of the BASELINE IMPEDANCE -switch and check that appropriate RESP waveform and RR value are shown again within 30 seconds.
Service check 6.2.7 Temperature measurement Enter the ESTP : STP service menu: Modules - ESTP : STP Check that the “Timeouts”, “Bad checksums” and “Bad c-s by mod” values are not increasing faster than by 50 per second. Check also that the STP board memories have passed the internal memory test, i.e.
Service check 6.2.10 Non Invasive Blood Pressure measurement Enter the NIBP module service menu: Monitor Setup - Install/Service (Password 16-4-34) - Service (Password 26-23-8) - Modules - NIBP Check that the “Timeouts”, “Bad checksums” and “Bad c-s by mod” values are not increasing faster than by 50 per second.
Page 50
Datex-Ohmeda S/5 Light Monitor the displayed pressure value in order to determine the real pressure. Check the watchdog timer activation pressure. Select Pneumatics from the NIBP service menu. Keep the pressure manometer connected to the cuff connector. Pump up the pressure very slowly and note the value on the manometer when you hear a signal from the loudspeaker.
Service check that the message “Cuff occlusion” appears on the screen within 30 seconds. Check that automatic inflation limits are in use: Menu – NIBP & Inv.Press - NIBP Setup&Alarm - Inflation Limits – Auto Previous Menu Connect the cuff onto your arm, highlight START VEN.STASIS in the NIBP & Inv.Press menu and press the ComWheel.
Datex-Ohmeda S/5 Light Monitor 6.2.12 For Light Monitor Perform electrical safety check and leakage current test. Check that the monitor functions normally after the performed electrical safety check. Clean the monitor with suitable detergent. • Fill in all necessary documents.
Adjustments and calibrations ADJUSTMENTS AND CALIBRATIONS 7.1 Pressure safety level detection “OFFSET” Check from the service view (Modules - NIBP) that AD0 = 0 ±1. If it is not: remove the back panel and adjust the trimmer located NIBP boards upper edge AD0 to zero. Pump up the pressure very slowly and check that the value of AD5 changes from negative value to positive value under pressure 11..15 mmHg.
Datex-Ohmeda S/5 Light Monitor Pump the following pressures to manometer and check the difference between the manometer and monitor pressure display: Table 2 NIBP calibration check pressures Pressure Max. error Example 0 mmHg ±9 mmHg (=zero offset) 100 + zero offset ±2 mmHg 98 ±2...
Adjustments and calibrations • Connect an external mercury manometer with pump to device through the both tubes of the hose - both transducers B1 and B2 must be calibrated simultaneously. Pump up to a pressure about 200 mmHg according to the manometer. Calibration is possible in the range 150 to 300 mmHg.
Datex-Ohmeda S/5 Light Monitor Press the ComWheel two times. Press in the protect button at the bottom of the module and choose ON in protect mode. Release the button. 7.4 Invasive pressure calibration Calibrate invasive pressure when the pressure transducer (probe) is replaced with a different type of transducer, and when LM-STP board is replaced.
General troubleshooting GENERAL TROUBLESHOOTING 8.1 Troubleshooting flowchart Monitor does not start Remove mains cable and batteries and switch monitor OFF Connect mains cable back and switch monitor Over voltage or over current protector triggered Fan is Check mains MAINS led running? cable lit?
Datex-Ohmeda S/5 Light Monitor SERVICE MENU 9.1 SW version / unit ID Software version and Unit ID are shown in the Service View menu. Menu Press key. Enter Monitor Setup - Install/Service (password 16-4-34) - Service View (password 26-23-8). Software strings have the following format: XXXX SW PPPPPP-N.N YYYY-MM-DD...
Service menu 9.2 Language selection Language can be selected in the Monitor menu. Menu Press key. Enter Monitor Setup - Install/Service (password 16-4-34) - Service View (password 26-23-8) - Monitor. 9.3 Monitor configuration Monitor configuration can be selected in the Monitor menu. Menu Press key.
Datex-Ohmeda S/5 Light Monitor GENERAL SPECIFICATIONS Safety standards Designed to meet IEC 60601-1 and CAN/CSA C22.2 No 601.1 10.1 Parameter specifications 10.1.1 ECG Measured leads: I, II or III Frequency response: 0.5...30 Hz (-3 dB, with 50 Hz reject filter) 0.5...40 Hz (-3 dB, with 60 Hz reject filter)
2, 5, 10, 20, 50 mod % or automatic scaling Accuracy is based on deep hypoxia studies using Datex-Ohmeda Finger Sensors on volunteered subjects. Arterial blood samples have been analyzed by a Radiometer OSM Co-oximeter. Refer to Datex-Ohmeda Sat Sensor directions for specific SpO accuracy.
General specifications 10.4.2 Trends Graphical: 2, 4, 12 and 24 h Numerical: all parameters, sampled every 5 min and after NIBP measurement Trend cursor: Both in graphical and numerical trends 10.4.3 Alarms • Adjustable high and low alarms for HR, Resp, NIBP, SpO2, Temp, CO2 (Et/Fi) and Invasive Pressures Alarm system classified according to priority;...
Datex-Ohmeda S/5 Light Monitor 10.8 Light External Battery Charger (optional) Mains voltage 100...240 VAC Max. input power 90 VA Mains frequency 50...60 Hz Charge current 1.7 A, ± 10 % Discharge current 0.6 A, ± 10 % Typical charging time 1 h per battery / 1.8 Ah...
FUNCTIONAL CHECK FORM Datex Ohmeda S/5 Light Monitor Customer Service Service engineer Date Monitor configuration Monitor model: F-LM1 Model with 2 InvBP measurements, F-LMP1 Parameter measurement options Data collecting and data management options N-LCM Mainstream CO N-LREC Recorder N-LNET Network...
Page 70
Datex-Ohmeda S/5 Light Monitor Performance checks N.A. Fail 1. Monitor screen configuration 2. SpO check 3. InvBP simulation 4. SpO : No probe and Probe off messages 5. SpO value (95-99 %) and HR value 6. SpO : No probe message 7.
SERVICE CHECK FORM Datex Ohmeda S/5 Light Monitor Customer Service Service engineer Date Monitor configuration Monitor model: F-LM1 Model with 2 InvBP measurements, F-LMP1 Parameter measurement options Data collecting and data management options N-LCM Mainstream CO N-LREC Recorder N-LNET Network...
Page 72
Datex-Ohmeda S/5 Light Monitor N.A. Fail 6.2.11 Mainstream CO measurement 6.2.12 For Light Monitor Electrical safety check and leakage current test Monitor functions after electrical safety check and leakage current test Cleaning Notes Signature B-2(2) Document No. M1051339...
APPENDIX C, EMC Guidance EMC GUIDANCE Guidance and manufacturer’s declaration – electromagnetic emissions Guidance and manufacturer’s declaration – electromagnetic emissions The S/5 Light Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the S/5 Light Monitor should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - guidance...
Page 74
Datex-Ohmeda S/5 Light Monitor Guidance and manufacturer’s declaration – electromagnetic immunity Guidance and manufacturer’s declaration – electromagnetic immunity The S/5 Light Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the S/5 Light Monitor should assure that it is used in such an environment.
Page 75
APPENDIX C, EMC Guidance Guidance and manufacturer’s declaration – electromagnetic immunity Guidance and manufacturer’s declaration – electromagnetic immunity The S/5 Light Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the S/5 Light Monitor should assure that it is used in such an environment. Immunity test IEC 60601 test Compliance...
Page 76
Datex-Ohmeda S/5 Light Monitor Recommended separation distances between portable and mobile RF communications equipment and the S/5 Light Monitor. Recommended separation distances between portable and mobile RF communications equipment and the S/5 Light Monitor. The S/5 Light Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Datex-Ohmeda S/5 Light Monitor Planned Maintenance Instructions All specifications are subject to change without notice. Document No. 8001934-2 September 2005 Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel. +1-608-221-1551...
Page 79
Table of contents TABLE OF CONTENTS S/5 Light Monitor Planned Maintenance Instructions TABLE OF CONTENTS PLANNED MAINTENANCE Introduction............................1 Recommended tools and accessories....................1 Recommended parts ........................... 1 INSTRUCTIONS Visual inspection ..........................2 General functional inspection....................... 3 Screen setup ............................4 Hemodynamic tests ..........................
Planned maintenance 1. PLANNED MAINTENANCE Introduction These instructions include procedures for planned maintenance (PM) for Datex-Ohmeda S/5 Light Monitor. The planned maintenance is recommended to be performed once a year after the monitor installation. The instructions are planned for the maximum functional configuration. Perform the procedures in the ascending order and skip the items that do not correspond to the monitor configuration.
Datex-Ohmeda S/5 Light Monitor 2. INSTRUCTIONS • Switch the monitor to standby. • Disconnect all the connection cables from the monitor’s rear panel. 2.1 Visual inspection 1. Make visual inspection for all external parts of the monitor. Check e.g. that the power cord receptacle is intact in the power adapter.
Planned maintenance Shielding Flaps (6 pcs) Figure 3 LM1-STP board EMC covers 2.2 General functional inspection 6. Connect the power cord to the power adapter and the power adapter cable to the Light Monitor. Check that the green LED turns on in the power adapter and the LED on the lower right-hand corner of the monitor turns on or starts flashing.
Datex-Ohmeda S/5 Light Monitor 8. Enter the keyboard menu and select Dummy Press. Monitor Setup - Install/Service (Password 16-4-34) - Service (Password 26-23-8) –Keyboard – Dummy Press Press the function keys one by one. Check that each key generates a sound from the loudspeaker and the corresponding text.
Page 85
Planned maintenance Non invasive blood pressure measurement tests 15. Enter the NIBP service menu and check that the NIBP board memories have passed the internal memory test, i.e. “RAM”, “ROM” and “EEPROM” all show OK. 16. Check the NIBP tubing system for leakages in Calibrations - Active Leak Test menu. The pressure must not drop more than 6 mmHg per one minute.
Datex-Ohmeda S/5 Light Monitor Invasive blood pressure measurement tests 26. Check the InvBP channels with a patient simulator. The values and waveforms should correspond to the simulator settings. Note: In case you evaluate the measurement accuracy with the simulator, add simulator's accuracy specification to the one of the monitor.
Planned maintenance Check that the monitor connects to the Datex-Ohmeda Network, i.e. the network symbol is displayed on the upper right-hand corner of the screen. A message regarding the connection to Datex-Ohmeda Central should appear in the message field on the screen.
Planned maintenance form PLANNED MAINTENANCE FORM Datex Ohmeda S/5™ Light Monitor Customer Service Service engineer Date Monitor configuration F-LM1 Model with 2 InvBP measurements, F-LMP1 Monitor model: Parameter measurement options Data collecting and data management options N-LCM Mainstream CO N-LREC...
Page 92
Datex-Ohmeda S/5 Light Monitor N.A. Fail 2.6. Recorder unit test 2.7. Network test 2.8. DataCard test 2.9. Service Log 2.10. Electric safety check Notes Used spare parts Fan filter Recorder paper Signature A-2(2) Document No. 8001934-2...
Functional description FUNCTIONAL DESCRIPTION 1.1 General The CPU board performs the central data processing of the Light Monitor. The operating principle of the CPU board is illustrated in figure 1. The CPU board is based around an embedded 386EX processor. Additionally, the CPU board includes a real time clock, battery back-up RAM, serial controller, audio generator, D/A converter and display controller connected to the external processor bus.
Datex-Ohmeda S/5 Light Monitor 1.1.1 386EX processor (D8) The 386EX processor is based on a 386SX core. Additionally the processor includes the following features: Programmable chip select lines • Boot block flash • Code flash • DRAM • Real time clock •...
Functional description CAUTION Discard lithium battery according to local regulations. 1.1.3 Serial controller, UART (D19) The serial controller is a four channel UART used for serial communication with the following units: • Serial I/O • CO 2 Unit • Power supply unit 1.1.4 Audio generator (D6, D14, N4) The audio frequency signal is generated by a timer integrated to the processor.
Functional description 1.2.4 ISA bus connector (X4) Signal Description Signal Description 1 = Access cycle, +5VA +5V Analog 0 = Transfer cycle BALE Addr Latch Enable GND A Analog gnd IOCHRDY Wait state needed +5V Supply IOCS16# Accessed recource is 16bits +5V Supply MEMCS16# Accessed recource is 16bits...
Datex-Ohmeda S/5 Light Monitor Signal Description Signal Description First 1Mega address Ground (memory & I/O) -”- IDDO Ethernet ID data output -”- IDPE Ethernet ID prg enable -”- Ground -”- IRQ7 Interrupt -”- SD15 Data Ground SD14 Data SD13 Data...
Functional description 1.2.6 Power supply board connector (X6) Signal Description Signal Description RSTOUT O,TTL CPU reset out RESETC# I,TTL CPU RESET NCALL NurseCall CMOS TXDPWR O,TTL Data from CPU DSYNC DefibSync -pulse CMOS RXDPWR I,TTL Data to CPU Module bus ANOUT0 O,±5V Analog output, channel 0...
Datex-Ohmeda S/5 Light Monitor 1.2.7 Recorder connector (X7) Signal Description Signal Description Not connected Not connected RXDREC I,TTL Serial receive -”- Not connected -”- +5VREC Voltage supply for the recorder -”- +5VREC Voltage supply for the recorder -”- GNDREC Recorder ground -”-...
Functional description 1.2.8 Module bus connector (X8) The NIBP-, ECG-, and STP-units are connected to module bus connector. Signal Description +12VMOD +12V module voltage +12VMOD +12V module voltage GNDMOD Ground Reserved Reserved Reserved O, CMOS Ground ESTPCLK ECG/STP clock GNDMOD Ground +5VMOD +5V module voltage...
Datex-Ohmeda S/5 Light Monitor 1.2.9 Display connector (X9) Signal Description Scan control TEST Test mode +3.3V Supply Voltage +3.3V Supply Voltage Data enable Ground Blue data signal Blue data signal Blue data signal Blue data signal Blue data signal Blue data signal...
Service procedures SERVICE PROCEDURES 2.1 General service information Field service of the CPU is limited to replacing a faulty circuit board or a real time clock battery. The boards should be returned to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the monitor serial number, full type designation and a detailed fault description.
Datex-Ohmeda S/5 Light Monitor 2.3 Replacing the CPU board Replace the new CPU board and make sure that all connections are intact correctly. Replace the rear cover and the chassis cover Enter Monitor Setup - Install/Service (password 16-4-34) - Service View (password 26-23-8) –...
Troubleshooting TROUBLESHOOTING 3.1 Troubleshooting flowchart Monitor doesn't start Check external power supply unit, optional Is the fan running? battery module and power supply board. Reset output signal in X1/a1 Failure in CPU-board. changes between 0V Change CPU-board. and +5V? Display or display Reset output control doesn't work.
Datex-Ohmeda S/5 Light Monitor 3.2 Troubleshooting chart TROUBLE CAUSE TREATMENT Real time clock adjustment has been lost Normal drift Readjust time Battery worn out Replace battery, adjust real time clock and make user configurations. User defaults have been lost Battery worn out Replace battery, adjust real time clock and make user configurations.
Page 115
Table of contents TABLE OF CONTENTS POWER SUPPLY UNIT TABLE OF CONTENTS TABLE OF FIGURES FUNCTIONAL DESCRIPTION General..............................1 Voltages .............................. 1 1.2.1 Input power sources........................1 1.2.2 Module voltages .......................... 3 Connectors on rear panel........................4 1.3.1 External power supply, X1......................5 1.3.2 Serial interface, X2........................
Functional description FUNCTIONAL DESCRIPTION 1.1 General The power supply board distributes power to all units in the Light Monitor. The power source is either an external power adapter, or one of the two rechargeable batteries (optional). If the monitor is used with mains power and the power cord is disconnected, operation continues with batteries without interruption.
Page 118
Datex-Ohmeda S/5 Light Monitor FA N V EXT BA TT #1 BA TT #2 Pow er/Charge P12VD Sw itches BA CKUP BA TTERY P5VD DC/DC P5V C Over V oltage Protection P3V3D DC/DC Sw itch Logic and P5V C V oltage regul.
Functional description 1.2.2 Module voltages +3.3V and +5V module voltages are made from 12 V monitor voltage by switch mode power supply. +12VDC voltage fluctuates according the table module voltages since battery voltage is unregulated. Backup battery Figure 2 Power supply block diagram Table 1 Module voltages Nom.
Functional description 1.3.1 External power supply, X1 External power supply connector (front view) External power supply connector pin assignments Signal Description Monitor voltage Ground Charge voltage, current limited Ground 1.3.2 Serial interface, X2 The connector provides a RS-232 serial communication link with handshaking. Pin out of the D9 male connector (front view) RS-232 Serial communication connector pin assignments Signal...
Datex-Ohmeda S/5 Light Monitor 1.3.3 Feature connector, X3 The connector provides module bus output with reset, two analog outputs, defib sync output and Nurse Call -signal with floating relay contact. Pin out of the high density D15 male connector (front view)
Functional description 1.3.4 Network 1.3.4.1 Ethernet interface, X5 The ethernet interface meets IEEE 802.3 specifications (10BASE-T) with hospital grade approved power and data transformers. Network connector (front view) Ethernet connector pin assignments Signal Description Tx + Transmit Data Tx - Transmit Data Rx + Receive Data...
Page 124
Datex-Ohmeda S/5 Light Monitor 1.3.4.2 Connector for network location ID plug, X6 The coding element interface is used for bedside address coding for the ethernet. The element is plugged in to the female D-connector of 9-pins. Pin out of the D9 female connector (front view)
Functional description 1.3.5 VGA connector, X7 Pin out of the high density D15 female connector (front view) The pin assignments of the VGA-connector Signal Green Blue digital gnd analog gnd (red) analog gnd (green) analog gnd (blue) digital gnd (logic) Horizontal sync.
Datex-Ohmeda S/5 Light Monitor 1.4 Defibrillator synchronisation output Defibrillator synchronisation connector Defibrillator synchronisation connector pin assignments Pin I/O Signal Signal Description Signal levels GNCOEQ Signal ground DSYNC DSYNC Defib. Synchronization output CMOS, active high, I out max 2.5 mA ANOUT0 ECG_OUT -5 V...+5 V...
Service procedures SERVICE PROCEDURES 2.1 General service information Field service for the power unit is limited to replacing a faulty circuit board. The board should be returned to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the monitor serial number, full type designation, and a detailed fault description.
Troubleshooting TROUBLESHOOTING Before troubleshooting the power supply board, please refer to the general troubleshooting procedures described in part I. 3.1 Troubleshooting flowchart Monitor doesn't start, but fan is running Check connectors, Measure output voltages of R55, R56 and X7/A 25 U(R55) >...
Service menu SERVICE MENU 4.1 Power menu Enter the Power menu according to the following procedures. Menu Press the key. Enter Monitor Setup - Install/Service (password 16-4-34) - Service View (password 26-23-8) - Modules - Power. Factory Test Battery related data (see next page). Power Service Batt1 type indicates battery 1 cell type.
Datex-Ohmeda S/5 Light Monitor 4.1.2 Factory test menu Factory Test Data Pow src: Pow src indicates present power source: EXT - external power supply, BATT 1 - battery 1, BATT 2 - battery 2. Batt 1/2 exists Batt 1/2 exists indicates battery existence: 0 - battery removed, 1 - battery inserted.
Page 133
Service menu Batt. last cap Batt. last cap indicates last measured full capacity of the battery. The capacity is calculated from a battery that is discharged from full to empty. Batt. used cap indicates capacity discharged from the battery. Remaining capacity: last cap - Batt.
Page 135
LM-NIBP board LM-ECG board LM-STP board Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel. +1-608-221-1551 FIN-00031 GE, FINLAND Tel: +358 10 39411 Fax +1-608-222-9147 Fax : +358 9 146 3310 www.datex-ohmeda.com www.gehealthcare.com...
Page 137
Table of contents TABLE OF CONTENTS Hemodynamics Boards TABLE OF CONTENTS TABLE OF FIGURES INTRODUCTION FUNCTIONAL DESCRIPTION Measurement principle......................... 3 2.1.1 NIBP ............................3 2.1.2 ECG............................. 3 2.1.3 Pulse oximetry ..........................3 2.1.4 Temperature ..........................6 2.1.5 Invasive blood pressure........................ 6 2.1.6 Respiration..........................
Introduction INTRODUCTION The LM-NIBP, LM-ECG, and LM-STP boards are printed circuit boards that provide general hemodynamic parameters as listed below. Letters in the board name stand for: NIBP = Non-Invasive Blood Pressure ECG = Electrocardiography (+ Impedance Respiration) S = Pulse oximetry T = Temperature P = Invasive Blood Pressure Document No.
Functional description FUNCTIONAL DESCRIPTION 2.1 Measurement principle 2.1.1 NIBP NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure. The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is inflated with a pressure slightly higher than the presumed systolic pressure, and deflated at a speed based on the patient's pulse, collecting data from the oscillations caused by the pulsating artery.
Page 142
The oxygen saturation percentage SpO 2 measured by Datex-Ohmeda module is calibrated against the functional saturation SaO 2 func. The advantage of this method is that the accuracy of SpO 2 measurement relative to SaO 2 func can be maintained even at rather high concentrations of carboxyhemoglobin in blood.
Page 143
Functional description Intensity of transmitted I max (DC-component) light I max AC-component I min Variable absorption Transmitted due to pulse added light volume of arterial blood Arterial blood Venous blood Tissue Time No pulsation Pulsatile blood Incident light SpO sensor connector SpO sensor cable Emitter IRED...
Applying a constant current through the resistor and measuring the voltage that is generated across it. In Datex-Ohmeda solution the two methods are combined in a form of a voltage divider. The NTC- resistor is connected in series with a normal resistor and a constant voltage is applied across them.
Functional description 2.2 Main components 2.2.1 Side panel NIBP ECG + Resp Signal Figure 2 Side panel connectors Mainstream CO connector Connectors for invasive pressure lines NIBP hose connector Pulse oximetry sensor connector ECG trunk cable connector, also used for impedance respiration measurement Temperature probe connector Signal out connector Document No.
Datex-Ohmeda S/5 Light Monitor 2.2.2 LM-NIBP board to/from CPU Figure 3 LM-NIBP board functional block diagram Pressure transducers The NIBP board contains two pressure transducers. They are of piezoresistive type. One is used for measuring the pressure of the blood pressure cuff and the pressure fluctuations caused by arterial wall movement (B1).
Page 147
Functional description Signal processing Two signals from the pressure transducers are amplified and sent to A/D converter. After the converter, digitized signals are sent to microprocessor for data processing. Before the converter, one of the signals is used to adjust the offset to the pressure safety level. The NIBP module is controlled with 80C51FA microprocessor at 16 MHz oscillator frequency.
Page 148
Datex-Ohmeda S/5 Light Monitor Valves Exhaust valves are used for emptying the cuff and the joining chamber after the measurement. Exhaust valve 1 is also used as safety valve in infant mode. Valve opens at 165 mmHg. Exhaust valve 2 is also used as safety valve in adult mode and opens at 320 mmHg.
Functional description 2.2.3 LM-ECG board Patient signals are connected to overload protection circuits (resistors and gas-filled surge arresters) and analog switches to instrumentation amplifiers. Then the signals are amplified by 480 and limited by slew rate. Then they are A/D-converted, analyzed and transferred to module bus in digital form.
Page 150
Datex-Ohmeda S/5 Light Monitor Noise and pacemaker are detected by a slew rate detector. Lower frequency is determined by a high pass (HP) filter 0.5 Hz (monitor bandwidth). ECG filtering The Light Monitor has two ECG filtering modes: MONITORING 0.5 - 30 Hz (-3 dB, with 50 Hz reject filter) 0.5 - 40 Hz (-3 dB, with 60 Hz reject filter)
Page 151
Functional description Microprocessor section Microprocessor contains RAM and EPROM memories. The processor uses external EEPROM memory for storing intendifying data. The microprocessor's internal 8-channel A/D-converter converts the ECG-signals to digital form. Serial communication Communication with the LM module bus is made through RXD and TXD signals. See the principle of serial communication in next figure.
Datex-Ohmeda S/5 Light Monitor Isolated section The patient isolation of ECG is 5 kV in ECG board. WARNING Do not touch any part of the monitor on patient connections during defibrillation procedure. Power supply section See the power supply section in LM-STP board description.
Page 153
Functional description Microprocessor unit As processor, Intel's 80C196KC-16 is used. There are external memories, an 8-bit data bus, a 16 MHz oscillator, within the processor there is an open collector reset, and a watchdog timer. Two A/D-converters within the processor are used. The processor's internal UART communicates with the CPU board.
Page 154
Datex-Ohmeda S/5 Light Monitor Pressure transducer Instrumentation amplifier Vout Input Filter to external AD converter Figure 8 Pressure transducer working principle Pulse oximetry measurement section LED control signals The processor sends pulse width modulated signals, IRED intensity and RED intensity, which are converted to DC voltage and filtered.
Page 155
Functional description Red and infrared channel separation Red and infrared channels are separated from each other by switches. Operational amplifier functions as a buffer and after this infrared and red DC signals are sent to the processor. A capacitor separates AC signal from DC signals and the AC signals are sent to the processor after amplification.
Page 156
Datex-Ohmeda S/5 Light Monitor Isolated section Signal isolation There are two opto isolators to isolate serial communication between LM-STP and CPU board. Reset line is an open collector type with a pull-up resistor. With that signal the processor is able to use its internal watch-dog function.
Functional description 2.3 Connectors and signals 2.3.1 Module bus connectors Signal Description CPU/NIBP/ECG P12VD + 12 V module voltage P12VD + 12 V module voltage Ground Ground ESTPCLK ECG/STP power clock Ground P5VD + 5 V module voltage TXDM Data to modules RXDM Data from modules RSTM#...
Datex-Ohmeda S/5 Light Monitor 2.3.2 Patient panel connectors ECG connector (ECG) BLUE Pin No. Signal Right arm electrode (R) Left arm electrode (L) Right leg electrode (RL) Left leg electrode (F) Chest electrode (C) Cable shield Not connected 3/5 lead identification...
Functional description 2.3.3 Test points LM-ECG and LM-STP board There are voltage test pins both on the LM-STP and LM-ECG boards. Pins and voltages are as follows: LM-ECG pin 1 +5 V pin 2 +5 Vref pin 3 +12 V pin 4 pin 5 -12 V...
Datex-Ohmeda S/5 Light Monitor SERVICE PROCEDURES 3.1 General service information Field service of the printed circuit boards is limited to replacing the faulty printed circuit boards or mechanical parts. The printed circuit boards should be returned to GE Healthcare for repair.
Page 161
Service procedures Parts needed • NIBP pump filter (order code 57141) Planned maintenance is recommended to be performed according to the following instructions at least once a year and always after service or repair. Check internal parts: • screws are tightened properly •...
Page 162
Datex-Ohmeda S/5 Light Monitor Replace the NIBP pump filter (p/n 57141). • Switch the monitor on and wait until the monitoring screen appears. Configure the monitor screen so that all the needed parameters are shown, for example as follows: Menu...
Page 163
Service procedures ECG measurement Enter the main menu: Menu - Monitor Setup - Install/Service (password 16-4-34) - Service View (password 26-23-8) Take down the information regarding ECG, STP, and NIBP board software (SW) by selecting Scroll Vers and turning the ComWheel or choose Records Vers if the recorder is installed. Enter the ESTP : ECG service menu: Menu - Monitor Setup - Install/Service (password 16-4-34) - Service View...
Page 164
Datex-Ohmeda S/5 Light Monitor NOTE: When the ground lead (RL, right leg) is disconnected, all the electrodes should show OFF. Connect a 3-lead cable to the monitor. Connect the leads together (e.g. to a screwdriver). Select one lead (I, II or III) for display.
Page 165
Service procedures Set the pacemaker option off: ECG – PACE - WAVE Respiration measurement Check that the “Resp Available” and “RESP Measurement” show both ON in the ESTP:ECG service menu. Do the following settings: Menu - ECG & Resp - ECG Lead and use 3-lead ECG cable.
Page 166
Datex-Ohmeda S/5 Light Monitor Temperature measurement Configure the monitor waveform field 4 so that it shows pleth waveform: Menu - Monitor Setup - Waveform Fields - Field 4 Pleth Enter the ESTP : STP module menu: Menu - Monitor Setup - Install/Service (password 16-4-34) - Service View (password 26-23-8) –...
Page 167
Service procedures For F-LMP1 with the invasive blood pressure measurement In the ESTP Module menu: Check that the “Cable” and “Probe” for P1 show OFF in the Service Data field. Plug a cable with an invasive blood pressure transducer into the front panel connector P1 and check that the “Cable”...
Page 168
Datex-Ohmeda S/5 Light Monitor Disconnect the SpO 2 sensor and check that “Probe off” message appears. Non invasive blood pressure measurement Enter the NIBP module service menu: Menu - Monitor Setup - Install/Service (password 16-4-34) - Service View (password 26-23-8) – Modules - NIBP Check that the “Timeouts”, “Bad checksums”...
Page 169
Service procedures Calibration check. Recalibrate the NIBP measurement. Remember to set the calibration protection back on after the calibration. Disconnect the pressure manometer. Select Calibrations and then highlight Calibration Check. Press the ComWheel and take down the zero offset values for both pressure transducers, B1 and B2.
Page 170
Datex-Ohmeda S/5 Light Monitor Check the watchdog timer also in the infant mode by first selecting Test INFANT from the menu. The time from the infant test should fall within 60-70 seconds (counter stops + beep). Checking the safety valve feature.
Page 171
Service procedures Keep the pressure inside the cuff for about half a minute in order to see that the cuff is not Normal Screen leaking, then press the ComWheel again. Select Keep the cuff on your arm and perform one NIBP measurement. Check that the measurement gives a reasonable measuring result.
Datex-Ohmeda S/5 Light Monitor 3.3 Adjustments and calibrations NIBP Pressure safety level detection "OFFSET" Check from the service view (Modules -> NIBP) that AD0 = 0 ±1. If it is not: remove the back panel and adjust the trimmer located NIBP boards upper edge AD0 to zero. Pump up the pressure very slowly and check that the value of AD5 changes from negative value to positive value under pressure 11…15 mmHg.
Page 173
Service procedures Pressure Max. error Example ±9 mmHg (=zero offset) 0 mmHg 100 mmHg 100 + zero offset ±2 mmHg 98 ±2 200 + zero offset ±3 mmHg 198 ±2 200 mmHg If the error of pressure channel B1 is larger than specified above, the module should be recalibrated.
Page 174
Datex-Ohmeda S/5 Light Monitor Verify that both pressure values in the prompt field match the manometer reading. If not, adjust by turning the ComWheel. When the values of the pressure bar and the manometer are equal, press the ComWheel to confirm the calibration. The message "Calibrating" will appear in the NIBP digit field.
Page 175
Service procedures Press in the protect button at the bottom of the module and choose ON in protect mode. Release the button. Invasive pressure calibration Calibrate invasive pressure when the pressure transducer (probe) is replaced with a different type of transducer, and when the LM-STP board is replaced. Enter ESTP: the STP service menu.
Datex-Ohmeda S/5 Light Monitor TROUBLESHOOTING 4.1 Troubleshooting charts See also the User’s Reference Manual for more troubleshooting procedures. NIBP TROUBLE CAUSE TREATMENT No NIBP value displayed NIBP not selected on screen. Check monitor setup. Message ‘Artifacts’ Unsuccessful measurement due to patient movements or shivering.
Page 177
Troubleshooting TROUBLE CAUSE TREATMENT 1. Hose and/or cuff not connected. 1. Connect the hose and the cuff. Message ‘Cuff loose’ 2. Hose and cuff connected. Reason: • tighten the cuff • cuff loosely wrapped • replace cuff/hose • leakage in cuff or hose •...
Page 180
Datex-Ohmeda S/5 Light Monitor TROUBLE CAUSE TREATMENT 1. If no ECG waveform, check LEADS HR numerical display No heart rate available. OFF message and connect the shows '---' leads. 2. If ECG waveform exists, check heart rate source e.g. in the ECG+RESP: ECG Setup.
Page 181
Troubleshooting Pulse oximetry (SpO 2 ) TROUBLE CAUSE TREATMENT 1. No probe connected to the 1. Check probe connections. Message 'NO PROBE' monitor. 2. Probe faulty. 2. Change the probe. 1. Unsuitable site. 1. Try another site. Message 'PROBE OFF' though probe 2.
Page 182
Datex-Ohmeda S/5 Light Monitor Temperature TROUBLE CAUSE TREATMENT Message Faulty calibration. Perform calibration. If it does not help, check that front panel connector is 'TEMPERATURE ERROR' properly connected to LM-STP board. Check front panel, or LM-STP board. No temperature displayed Wrong type of probe.
Page 183
Troubleshooting TROUBLE CAUSE TREATMENT Message ‘Not zeroed’ Measurement on, channel not zeroed. Zero the channel. Message ‘Zeroing failed’ Unsuccessful zeroing of P1 /P2 Possibly due to pulsating pressure (number field). waveform. Open the transducer to air and zero the channel. Offset is >...
Page 184
Datex-Ohmeda S/5 Light Monitor Impedance respiration TROUBLE CAUSE TREATMENT No resp trace Waveform not selected on the screen. Enter the menu: Main Menu: Monitor Setup: Waveform fields and make adjustments Unacceptable resp waveform Poor electrode or poor electrode skin Electrodes from different contact.
Troubleshooting 4.1.2 NIBP Troubleshooting S T A R T P o ssib le fa u lt in N IBP Sw itch d e vice o ff a n d o n u n it a g a in F a u lt n o t in N IBP u n it D o e s fa u lt still a p p e a r? Ta ke m o d u le b u s Ye s...
Datex-Ohmeda S/5 Light Monitor 4.1.3 LM-STP and LM-ECG board troubleshooting Possibly faulty STP- or ECG-board Check the monitor type in service view: Modules: ESTP:STP Select the right Is it monitor configuration correct? in service view: Monitor: Configuration Remove module bus cable from the NIBP-board.
Page 187
Troubleshooting R e p la c e th e E C G R -b o a rd . D o e s th e fa u lt still a p p e a r? ye s R e p la c e th e S T P -b o a r d .
Service menu 5.1 NIBP service menu Service data Pressure shows measured pressure multiplied by 10. Zero shows pressure at auto zeroing multiplied by 10 and changes between +20 and -20 mmHg. Absolute pressure is the sum of Pressure and Zero. Protect handle indicates hardware protection for EEPROM memory.
Datex-Ohmeda S/5 Light Monitor 5.1.1 NIBP calibration menu Active leak test Wrap an adult cuff around a pipe and connect the cuff to the module. Select the active leak test (ON). The module automatically pumps a pressure of 260 mmHg into the cuff. Wait for several seconds until the pressure stabilizes.
Page 191
Service menu Calibration Calibration selection is available only when protection is OFF. NIBP calibration can be performed in the NIBP Service menu as follows: NOTE: Both channels B1 and B2 must be calibrated simultaneously. If Protection is ON, change it to OFF by first pressing the switch at the bottom of the monitor, which enables the Protection selection.
Datex-Ohmeda S/5 Light Monitor 5.1.2 NIBP safety valve menu ADULT/INFANT Selection of patient. Start test Start test is for starting and Stop test is for stopping the Safety Valve test. NOTE: Parameter values in Service Data are for reference only.
Service menu 5.1.3 NIBP pulse valve menu Start test Start test is for starting and Stop test is for stopping the test. Set valve Set Valve lets you adjust the opening of the pulse valve. NOTE: Parameter values in Service Data are for reference only. Pulse valve data for information on general items Pressure, Zero, Protect See NIBP service menu in chapter...
Datex-Ohmeda S/5 Light Monitor 5.1.4 NIBP pneumatics menu Set valve With Set Valve, the opening of the pulse valve is adjusted between 0 and 255 (0 for fully closed and 255 for fully open). First push the ComWheel, then turn it to adjust the value on the screen and finally push to set the value.
Service menu Watchdog BEEP Connect manometer to the front panel and pump pressure into the module. When the AD5 value changes from a negative to a positive value (at about 11 mmHg), a beep is heard. This is the watchdog threshold pressure. Beyond this pressure the watchdog is active and cuts pressures at about 2 min.
Datex-Ohmeda S/5 Light Monitor 5.2 ECG service menu Power freq Set power frequency (50 Hz/60 Hz). Service data Power freq and Cable type show the values chosen or detected. Permanent filter setting is: low 0.50 Hz. High 30 Hz depends on Power freq setting.
Service menu RAM indicates the state of the external RAM memory. ROM indicates whether the checksum at the EPROM is in accordance with the one the software has calculated. EEPROM indicates if the values stored in the permanent memory are valid. The state is either OK, Fail or ? (module not in place or a communication error).
Page 198
Datex-Ohmeda S/5 Light Monitor Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected. The Light Monitor starts counting these items at power up and resets to zero at power off. The nonzero values do not indicate a failure, but the continuous counting (more than 50 per second) indicates either a serial communication failure, or a board not in place.
Service menu 5.3.2 Calibration menu Protection Protection can be set to ON or OFF only when the protect button at the bottom of the LM-STP board is pressed. Set config, calibrate T1 Set Config and Calibrate T1 selections are available only when protect mode is off. Calibrate P1 and calibrate P2 Calibrate P1 and Calibrate P2 are available when pressure transducer is connected to the receptacle on the front panel.
APPENDIX A PLANNED MAINTENANCE CHECK FORM, LM-NIBP, LM-ECG, LM-STP BOARDS Customer Service Service engineer Date OK = Test OK N.A. = Test not applicable Fail = Test Failed All boards N.A. Fail N.A. Fail Internal parts External parts NIBP pump filter Parameter information Notes ECG measurement...
Page 204
Datex-Ohmeda Light Monitor Notes TEMP measurement N.A. Fail N.A. Fail 14. Communication and 15. Temperature probe memories detection 16. Calibration check 17. Temp test function Notes For F-LMP1 with the InvBP measurement 18. Cable and transducer 19. Calibration detection 20. Test with patient...
Page 205
APPENDIX A Infant 60-70 s 29. Safety valve feature Allowed range “Max press” ADULT 290-330 mmHg “2 s after stop” ADULT 290-330 mmHg “Max press” INFANT 154-165 mmHg “2 s after stop” INFANT 154-165 mmHg 30. Cuff related 31. Adult cuff detection messages 32.
The CO board is designed for use with Datex-Ohmeda sensors and cables. 1.2 Measurement principle measurement is based on the infra-red absorption characteristic of CO molecules.
Page 212
Datex-Ohmeda S/5 Light Monitor REFERENCE+ SAMPLE SAMPLE DARK TIME Figure 2 Photodetector voltage As the single IR beam encounters the three areas of the rotating chopper wheel and the airway sample, a signal is generated. A ratio of “reference + patient sample/patient sample” is used to determine the CO concentration in the patient’s respiratory gas.
Functional description 1.3 CO module The operating principle of the CO module is illustrated in figure 3. The system can be divided into the following functional parts: • Main Processor Board • Mainstream daughter board • Isolated Power Supply MAINSTREAM CO2 SENSOR CO2 MODULE MAINSTREAM STICK...
Datex-Ohmeda S/5 Light Monitor 1.3.3 Isolated power supply unit Please see the next slot (5) “Isolation Power Supply Unit for CO Unit”. 1.3.4 Serial communication The CO module, manufactured by Welch Allyn, is used in the Light Monitor for CO measurement.
Functional description 1.4 Internal connectors Connector for CO module (P105 on DuET board) Signal Function + 10 V Input Power Ground Data from CO board to Host Data from Host to CO board Mainstream flex circuit header (P302 on MS-Stick) Signal Function - 5 Volts...
Service procedures SERVICE PROCEDURES 2.1 General service information Field service for the CO board is limited to the replacement of faulty circuit boards or mechanical parts. The board should be returned to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the monitor serial number, full type designation, and a detailed fault description.
Fix the mainstream sensor to the CO calibration airway adapter. Feed 5.00 % CO gas (e.g. Datex-Ohmeda Quick Cal Calibration Gas, order No. 755580) to the free end of the silicon tube (included in the kit P/N 902305). Introduce a steady stream of gas until FiCO and EtCO readings are same and observe the reading.
Page 219
Service procedures Possible causes of error Calibration should be performed using a CO calibration gas of a known concentration. To validate the concentration of the calibration gas, check by using several different sensors. If they are all consistently high or low, there is a good chance that the gas sample is suspect and another gas source should be tried.
Datex-Ohmeda S/5 Light Monitor SERVICE MENU 3.1 CO module menu NOTE: Parameter values in every Data field are for reference only. Sensor data See chapter Sensor menu. 3.1.2 Pump data See chapter Pump menu. 3.1.3 Service data HW version Hardware version of a CO board.
Serial number of a sensor. sensor rev. Sensor rev. shows a revison of the sensor. Custom code must be 10 with Datex-Ohmeda MS sensor. Sensors by other manufacturers are not custom code possible and generate alarm ‘Use Datex-Ohmeda sensor’. temperature (C) Temperature shows the temperature of a sensor (°C). Temperature of a warm-up MS sensor should be 40...42 °C.
Datex-Ohmeda S/5 Light Monitor 3.1.3 Pump menu NOTE: Parameter values in every Data field are for reference only. Pump data Flow rate (ml/min) Flow rate shows adjusted flow rate which should be 150 ml/min. Pump ctrl value Pump ctrl value is comparable to supplied power to the pump.
Page 223
ISOLATION POWER SUPPLY UNIT FOR CO UNIT Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel. +1-608-221-1551 FIN-00031 GE, FINLAND Tel: +358 10 39411 Fax +1-608-222-9147 Fax : +358 9 146 3310 www.datex-ohmeda.com www.gehealthcare.com...
Page 225
Table of contents TABLE OF CONTENTS ISOLATION POWER SUPPLY UNIT FOR CO 2 UNIT TABLE OF CONTENTS FUNCTIONAL DESCRIPTION General..............................1 Connectors ............................2 SERVICE PROCEDURES General service information ........................3 TROUBLESHOOTING Troubleshooting flowchart........................4 Document No. 8000060-1...
Functional description FUNCTIONAL DESCRIPTION 1.1 General The power supply distributes power to CO unit providing 1.5 kV patient isolation. Input voltage range is 10.0...15.0 V. Output voltage is regulated to 10.0 V ±2.5 %. In the case of malfunction of feedback circuit, output voltage is limited to 11.0 V. Isolation power supply can be shut down for power saving by the host, if CO measurement is not used.
Datex-Ohmeda S/5 Light Monitor 1.2 Connectors unit connector (X1) Pin No. Signal Description WTPR# SideStream Water Trap present GNDF Signal ground GNDF Signal ground +10V +10V output voltage TXDF Data from CO unit (TTL-level) RXDF Data to CO unit (TTL-level) Host connector (X2) Pin No.
Troubleshooting SERVICE PROCEDURES 2.1 General service information Field service for the CO isolation power supply unit is limited to replacing a faulty circuit board. The boards should be returned to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the monitor serial number, full type designation, and a detailed fault description.
Datex-Ohmeda S/5 Light Monitor TROUBLESHOOTING Before troubleshooting the power supply unit, please refer to the general troubleshooting procedures described in part I. 3.1 Troubleshooting flowchart measurement doesn't operate Check that CO2 measurement is selected ON (MS or SS in CO2...
Page 231
Datex-Ohmeda Light Monitor Recorder, N-LREC Technical Reference Manual All specifications are subject to change without notice. Document No. 890201-5 July 2005 Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel. +1-608-221-1551...
Page 233
Table of contents TABLE OF CONTENTS S/5™ Light Monitor Recorder, N-LREC TABLE OF CONTENTS INTRODUCTION SPECIFICATIONS FUNCTIONAL DESCRIPTION General..............................3 Connectors ............................3 3.2.1 Rec-board connector, X1......................3 3.2.2 Rec-board connector, X2......................4 SERVICE PROCEDURES General service information ........................5 Planned maintenance ..........................
Recorder, N-LREC INTRODUCTION The built-in recorder module of the S/5 Light Monitor/LIght Solo, N-LREC, provides real time printing of waveform and numerical data, and printing up to 12 hours trend data. NOTE: Printings on thermal paper may be destroyed when exposed to light, heat, alcohol etc. Take a photocopy for the archive.
Recorder, N-LREC FUNCTIONAL DESCRIPTION 3.1 General The recorder board connects the recorder module to CPU-board. The recorder board and the CPU- board are connected together with a 16-pin ribbon cable. Communication with the host processor on the CPU-board takes place in TTL-level serial communication.
Recorder, N-LREC SERVICE PROCEDURES 4.1 General service information Field service for the recorder is limited to replacing a faulty circuit board or mechanical parts. Return the board to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the unit serial number, full type designation, and detailed fault description.
Datex-Ohmeda S/5 Light Monitor 4.3 Disassembly and reassembly Disassemble the recorder according to the following procedure. Turn the monitor off and remove the power adapter cable. Install the recorder and tighten the screws on the back wall of the paper slot.
Recorder, N-LREC TROUBLESHOOTING 5.1 Troubleshooting chart TROUBLE CAUSE TREATMENT Recorder will not start. Recorder board ribbon cable Check the cable. Replace if broken. necessary. No error messages shown. Recorder board loose. Check recorder board connections. Recorder faulty. Replace the recorder. Recorder works but nothing Active side of paper Turn the paper roll.
Datex-Ohmeda S/5 Light Monitor SERVICE MENU There is no service menu available for the recorder. Document No. 890201-5...
Page 243
Light Solo POWER ADAPTER, N-LPOW Technical Reference Manual All specifications are subject to change without notice. Document No. M1051340 September 2005 Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel. +1-608-221-1551...
Page 245
Table of contents TABLE OF CONTENTS POWER ADAPTER, N-LPOW TABLE OF CONTENTS TABLE OF FIGURES INTRODUCTION SPECIFICATIONS FUNCTIONAL DESCRIPTION General..............................3 Connectors ............................3 SERVICE PROCEDURES General service information ........................4 Disassembly and reassembly ....................... 4 4.2.1 Changing output cable ......................... 5 TROUBLESHOOTING Troubleshooting chart...........................
Page 246
Datex-Ohmeda S/5 Light Monitor and Light Solo Document No. M1051340...
Specifications INTRODUCTION The Datex-Ohmeda Light Monitor Power Adapter, N-LPOW is an AC/DC converter for converting the mains voltage from the 220-240/100-120 Vac to voltages used by the monitor +12.5/+18.5 Vdc. NOTE: Before use check that the voltage range is selected properly.
Datex-Ohmeda S/5 Light Monitor and Light Solo SPECIFICATIONS Dimensions (W × D × H) 115 × 175 × 55 mm / 4.5 × 6.9 × 2.2” Weight 0.9 kg AC voltage range and frequencies 220-240/100-120 V~, 50/60 Hz Allowed AC voltage fluctuations 100 V -10 %...120 V +10 %...
Functional description FUNCTIONAL DESCRIPTION 3.1 General The power adapter is an AC/DC converter which converts the mains voltage to +12.5 Vdc and +18.5 Vdc. +18.5 Vdc is a 600 mA constant current source for battery charging. The green LED in the secondary end indicates whether the power adapter is energized. Chassis and secondary ground are connected to the protective earth (PE).
Datex-Ohmeda S/5 Light Monitor and Light Solo SERVICE PROCEDURES 4.1 General service information GE Healthcare is always available for service advice. Please provide the unit serial number, full type designation, and detailed fault description. CAUTION Only trained personnel with appropriate equipment should perform the tests and repairs outlined in this section.
Service procedures 4.2.1 Changing output cable Disassemble the frame as described in previous chapter. Remove the solderings (if cable is soldered to PCA) of the old cable and release the strain relief from the end plate. Solder PCA connector which came with the new cable to the PCA as shown in figure 3. Place the new cable and strain relief on its position on the end plate and connect the cable connector to the PCA connector.
Datex-Ohmeda S/5 Light Monitor and Light Solo TROUBLESHOOTING 5.1 Troubleshooting chart TROUBLE CAUSE TREATMENT Green LED is off Mains not connected Check power cord Green LED is off Blown fuse Check fuses Document No. M1051340...
Service menu SERVICE MENU There is no service menu available for the power adapter. Document No. M1051340...
Page 254
Datex-Ohmeda S/5 Light Monitor and Light Solo Document No. M1051340...
Page 255
Light Solo BATTERY MODULE, F-LBAT Technical Reference Manual All specifications are subject to change without notice. Document No. 890200-5 July 2005 Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel. +1-608-221-1551...
Page 257
Table of contents TABLE OF CONTENTS BATTERY MODULE, F-LBAT TABLE OF CONTENTS INTRODUCTION SPECIFICATIONS Battery module ............................ 2 Battery..............................2 FUNCTIONAL DESCRIPTION General..............................3 Connectors ............................4 3.2.1 Battery module connector ......................4 3.2.2 Battery connector ........................4 SERVICE PROCEDURES General service information ........................
Page 258
Datex-Ohmeda S/5 Light Monitor and Light Solo Document No. 890200-5...
Battery Module, F-LBAT INTRODUCTION The Battery Module, F-LBAT can be used with S/5 Light Monitor and Light Solo monitors. F-LBAT includes two rechargeable batteries. The batteries can easily be removed and replaced. Document No. 890200-5...
Datex-Ohmeda S/5 Light Monitor and Light Solo SPECIFICATIONS 2.1 Battery module Dimensions (W × D × H) 290 × 150 × 40 mm / 11.5 × 5.9 × 1.6“ Weight 0.5 kg (without batteries) 2.2 Battery Related to p/n 887753.
Battery Module, F-LBAT FUNCTIONAL DESCRIPTION 3.1 General The batteries are located in a battery module. The batteries can be inserted in the battery module from the left side of the monitor. The battery module is designed for two batteries. When the monitor is battery driven with two batteries, one battery is always discharged first. Less charged battery is discharged first.
Battery Module, F-LBAT SERVICE PROCEDURES 4.1 General service information Field service for the battery module is limited to replacing a faulty module. The module should be returned to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the unit serial number, full type designation, and detailed fault description.
Datex-Ohmeda S/5 Light Monitor and Light Solo 4.3 Installing the battery module Install the battery module according to the following procedure. Turn the monitor off and remove the power adapter. Place the monitor front side down on the table, use soft material between the table and the monitor.
Page 265
Battery Module, F-LBAT On Light Solo Turn the monitor on. Wait for three minutes. Verify that the battery status is displayed and Batt In use no error messages appear. Full Disconnect the monitor from the mains power and verify that the monitor operates with both batteries.
Datex-Ohmeda S/5 Light Monitor and Light Solo TROUBLESHOOTING 5.1 Troubleshooting chart TROUBLE CAUSE TREATMENT Monitor will not start with batteries. Batteries are empty. Charge batteries or replace them with full batteries. Batt Fault error. See error Battery Module not properly Reassemble module properly.
Battery Module, F-LBAT 5.3 Battery error messages on Light Solo Table 1 Battery errors, error codes on display and in service view BCOM BDATA TEMPFLT CELLFLT Error n Service Page (on display) Document No. 890200-5...
Page 268
Datex-Ohmeda S/5 Light Monitor and Light Solo Table 2 Setting and resetting the error bits Error bit Description Setting the bit Resetting the bit BCOM Communication The bit is set if reading the test If test byte read is successful.
Battery Module, F-LBAT 5.4 Battery error messages on Light Monitor See explanation from Power Supply Unit (Part II/slot 2), 4 Service Menu, 4.1 Power Menu, Power Service. Document No. 890200-5...
Datex-Ohmeda S/5 Light Monitor and Light Solo SERVICE MENU Please refer to the service menu for the “Power supply unit”, in part II, the slot 2. Document No. 890200-5...
Page 271
Light Solo LIGHT EXTERNAL BATTERY CHARGER, N-LCHGR Technical Reference Manual All specifications are subject to change without notice. Document No. 896192-1 July 2005 Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel.
Page 272
Intended purpose Battery Charger is intended for stationary indoor use. Charger is intended for charging the S/5 Light Monitor and Light Solo monitors’ batteries only. Classification according to IEC 60601-1 − CLASS I EQUIPMENT according to the type of protection against electrical shock. −...
Page 273
Table of contents TABLE OF CONTENTS LIGHT EXTERNAL BATTERY CHARGER TABLE OF CONTENTS INTRODUCTION Related documents ..........................1 Symbols .............................. 1 1.2.1 Symbols on the charger........................ 1 1.2.2 Symbols on the battery......................... 2 TECHNICAL SPECIFICATIONS Environmental specifications ........................ 3 Electrical specifications ........................3 Mechanical specifications ........................
Page 274
Datex-Ohmeda S/5 Light Monitor and Light Solo Document No. 896192-1...
INTRODUCTION This manual provides technical description and instructions for the service, maintenance and troubleshooting for the Datex-Ohmeda Light External Battery Charger. 1.1 Related documents The User’s Reference Manual (order code M1037860) advises you on how to use and maintain the Datex-Ohmeda Light External Battery Charger.
Datex-Ohmeda S/5 Light Monitor and Light Solo 1.2.2 Symbols on the battery Attention, consult accompanying documents: Do not expose to fire. Do not short-circuit any terminals. Do not disassemble or mutilate. Use only with the Light Monitor Battery Module or Light External Battery Charger.
Datex-Ohmeda S/5 Light Monitor and Light Solo FUNCTIONAL DESCRIPTION 3.1 General The device is an external charger for the LM Battery. Charger uses discharge-before-charge procedure in order to refresh a battery to its full capacity. The typical charge time is about 1 h for an 1800 mAh battery.
External Battery Charger, N-LCHGR Figure 2 Back panel Dust filter Power inlet and fuses 3.3 Charging 3.3.1 General Charger is always energized when the mains power is connected. The charging process starts automatically when a battery is inserted to the charger if any faults have been observed in the charge initialization routine.
Datex-Ohmeda S/5 Light Monitor and Light Solo SERVICE PROCEDURES 4.1 General service information Field service for the charger is limited to changing fuses and PCBs. GE Healthcare is always available for service advice. Please provide the unit serial number, full type designation, and detailed fault description.
External Battery Charger, N-LCHGR TROUBLESHOOTING Problem Possible cause Solutions Mains power ON indicator led on the Primary fuses are burned. Check both primary fuses, front panel is not illuminated. T2AH/250 V (p/n 511200). Medical power board fuse is burned. Check the fuse, 2.5 A (p/n 511181).
Datex-Ohmeda S/5 Light Monitor and Light Solo SERVICE MENU There is no service menu available for the external battery charger. SPARE PARTS Item numbers refer to the exploded view. 7.1 Light external battery charger, N-LCHGR (revision 00 and 01) Item Item description Order No.
Page 284
Datex-Ohmeda S/5 Light Monitor and Light Solo Document No. 896192-1...
Page 285
DATACARD OPTION, N-LDATA (REV. 00) NETWORK OPTION, N-LNET (REV. 00) DATACARD AND NETWORK OPTION, N-LDNET (REV. 00, REV. 01) Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel. +1-608-221-1551 FIN-00031 GE, FINLAND...
Page 287
Table of contents TABLE OF CONTENTS DATACARD OPTION, N-LDATA (REV. 00) NETWORK OPTION, N-LNET (REV. 00) DATACARD AND NETWORK OPTION, N-LDNET (REV. 00, REV. 01) TABLE OF CONTENTS INTRODUCTION TECHNICAL SPECIFICATIONS FUNCTIONAL DESCRIPTION General..............................3 DataCard option........................... 3 3.2.1 PCMCIA host adapter ........................3 3.2.2 Voltage switching matrix.......................
Introduction INTRODUCTION DataCard option DataCard option (N-LDATA or N-LDNET) is used for storing patient physiological data. The DataCard with its previously recorded patient data can be transported to a new monitor with the patient enabling continuous data collection. The files created in DataCard are MS-DOS compatible.
Datex-Ohmeda S/5 Light Monitor TECHNICAL SPECIFICATIONS General Total power consumption: 300 mW maximum Connection to the CPU-board: ISA-bus, 100pin connector DataCard option PC Card socket: one socket, type II Data storage capacity: 2 days of continuous physiological data trends File system:...
Datex-Ohmeda S/5 Light Monitor 3.2.3 Single PCMCIA socket Single PCMCIA Socket meets PCMCIA type II. There is a push button for removing the DataCard. DataCard connector allows to insert/remove the DataCard, when a monitor is switched ON. 3.3 Network option 3.3.1 Network controller...
Functional description 3.3.7 Network status LEDs There are three status LEDs for network communication: • ACTIVITY: is asserted for 50ms whenever the DP83907 transmits or receives data. • COLLISION: is asserted for 50ms whenever the DP83907 detects a collision. • GOOD LINK: is on if the link integrity is good (i.e.
Datex-Ohmeda S/5 Light Monitor CONNECTORS AND SIGNALS 4.1 Network connector See Power Supply Unit chapter 1.3.4. 4.2 Coding element interface See Power Supply Unit chapter 1.3.4.2. 4.3 ISA bus connector (X2) See CPU chapter 1.2.4. Document No. 8000060-1...
Connectors and signals 4.4 DataCard connector (X4) Signal I/O Description Signal Description Ground Ground Data CD1# Card Detect Data CD11 Data Data CD12 Data Data CD13 Data Data CD14 Data CE1# Card Enable CD15 Data CA10 Address CE2# Card Enable Output Enable CA11 Address...
Datex-Ohmeda S/5 Light Monitor SERVICE PROCEDURES Due to the nature of the network board, field service is limited to troubleshooting. The circuit boards should be returned to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the unit serial number, full type designation, and a detailed fault description.
Troubleshooting TROUBLESHOOTING 6.1 Troubleshooting charts Symptom at the monitor end Problem at Explanation/correction Monitor does not connect to Monitor-Network cable Cable not properly connected to wallplate network. or to monitor. Monitor connects to network, but disconnects unexpectedly ("Network xxxx down" message on the monitor screen.
Service view 7.1 Network menu Network Network displays network related service data. Please see the chart below. Value Usage Notes Received packets Total number of received packets (Statistics In/Packets) since last cold start. Transmitted packets Total number of transmitted packets since last cold start. (Statistics Out/Packets) Received bytes Total number of received bytes...
Datex-Ohmeda S/5 Light Monitor Location ID ID plug Must equal to ID plug Address Monitor’s ethernet address Connections Names of DEICs connected DMA packets *) Total number of DMAs to/from *) not seen on figure above controller Lost IRQ packets *) Number of lost IRQ packets 7.2 Computer interface menu...
Service view EARLIER REVISIONS PC/E-Board, 891437 Rev. 0 Functionally same as 891437 Rev. 1. PC-Board, 894697 There are no earlier revisions of PC-Board E-Board, 894698 There are no earlier revisions of PC-Board Document No. 8000060-1...
Page 303
UPGRADE KITS, SPARE PARTS AND EXPLODED VIEWS All specifications are subject to change without notice. Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel. +1-608-221-1551 FIN-00031 GE, FINLAND Tel: +358 10 39411...
Page 305
UPGRADE KITS, SPARE PARTS AND EXPLODED VIEWS TABLE OF CONTENTS Upgrade kits Installation Guides ..........................1 Spare Parts Replacing the LCD Backlight ......................... 2 Light Monitor, F-LM1 and F-LMP1......................3 2.2.1 Front cover and display unit......................3 2.2.2 Frame............................4 2.2.3 Frame, parameter unit........................
Page 306
Datex-Ohmeda S/5 Light Monitor This page intentionally left blank. Document No. M1051358...
Measured and entered patient data can be viewed on the IC. − Measured patient data and alarms can be viewed on other Datex-Ohmeda S/5 monitoring systems, but not vice versa. DataCard & Network Option Includes both options described above. N-LDNET...
Datex-Ohmeda S/5 Light Monitor SPARE PARTS Item numbers refer to the exploded view beside/below the table. NOTE: Accessories are listed in the Patient Monitor Supplies and Accessories Catalogue. a) means that the part is recommended for stock 2.1 Replacing the LCD Backlight In normal circumstances it is very difficult to keep the LCD Display component and the LCD Display shield free of dust, when they are detached from the monitor frame.
Spare parts 2.2.5 Back cover unit (896908) Item Item description Order No Shaft bar 886728 a) Rear enclosure 896690 Handle 896688 Lock ring 63900 a) Interlock 888932 a) Document No. M1051358...
Datex-Ohmeda S/5 Light Monitor 2.2.9 Chassis Item Item description Order No CPU Board for F-LM1-03 M1038993-S Battery for SRAM/Timekeeper 197230 Power Supply Board 891120 Board holder, F-LM1 M1053889-S Isolation plate 887992 Cross cyl.-head screw M3x6 tufflock (9 pcs) 617210 a) The part is recommended for stock...
Spare parts 2.2.12 NIBP unit Item Item description Order No NIBP board, LM1 890658 a) Pump unit 892518 a) Screw cross rec. cylind. h. stzn, M3x25 61742 Pneumatic unit 892517 a) 628712 PT screw M3×8 NIBP pump filter, LM1 57142 a) Pressure sensor 16539 Clip for pressure sensor...
Datex-Ohmeda S/5 Light Monitor RIBBON CABLES AND CONNECTORS Parameter Unit 892515 InvBP Input Board 891258 STP Board 890676 Input Board 891255 MS-receptacle TEMP cable 891547 893360 ECGR Board ECG Input Board CO2 Unit 890673 891544 N-LCSM 894752 (MS) 892526 CO2 Power...
Page 327
APPENDICES APPENDICES Installation Guides for N-LCM, N-LDNET, N-LDATA and N-LNET Method of Replacing the Lamp Unit (LCD Backlight)
Page 328
Datex-Ohmeda S/5 Light Monitor This page intentionally left blank.
Page 329
Datex-Ohmeda Optional N-LCM Upgrade for Light Monitors, F-LM1 and F-LMP1 Installation Guide All specifications are subject to change without notice. Document No. 896551-2 November 2000 Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel.
Page 331
Table of contents TABLE OF CONTENTS Optional N-LCM Upgrade for S/5™ Light Monitors, F-LM1 and F-LMP1 TABLE OF CONTENTS TABLE OF FIGURES INTRODUCTION Installation time ........................... 1 Tools needed ............................1 Parts needed ............................1 CONFIGURATION AND TESTS Mainstream CO configuration and test ....................
Page 333
Optional N-LCM Upgrade for Light Monitors, F-LM1 and F-LMP1 INTRODUCTION This guide describes how to install the mainstream CO measurement upgrade set for Light Monitors, F-LM1 and F-LMP1. The mainstream CO measurement upgrade set provides mainstream CO measurement capability. 1.1 Installation time Approximately one to one and a half hour.
Page 334
Datex-Ohmeda S/5 Light Monitor CONFIGURATION AND TESTS 2.1 Mainstream CO configuration and test Monitor Setup Enter – Install/Service (password 16-4-34) – Service View (password 26- 23-8) – Monitor – Configuration. Choose the correct set configuration related to the monitor hardware configuration. Perform factory reset.
Page 335
Optional N-LCM Upgrade for Light Monitors, F-LM1 and F-LMP1 CO2 hardware assembly 893373 Flex cable 902401 Figure 1 Parts The CO2 hardware is installed on top of the recorder housing. The NESTPR-module has to be removed for cabling to the patient connector plate.
Page 336
Datex-Ohmeda S/5 Light Monitor Install the module using the bushings and screws. Put the EMI filter on its holder. 640445 640445 617210 Figure 3 Installing the module On the flex cable are marks where to bend the cable Figure 4 Flex cable Document No.
Page 337
Optional N-LCM Upgrade for Light Monitors, F-LM1 and F-LMP1 Fasten the CO2 connector with the two screws provided in the kit 628703 Figure 5 connector Put the flex cable according to the picture Figure 6 Flex cable position Document No. 896551-2...
Page 338
Datex-Ohmeda S/5 Light Monitor Put the flex cable through the EMI filter. NOTE: Work very carefully with the flex cable which breaks easily Figure 7 Flex cable and the EMI filter Pull out the locking piece before pushing the flex cable into the connector.
Page 339
Optional N-LCM Upgrade for Light Monitors, F-LM1 and F-LMP1 Test the monitor using Main Stream sensor and Calibration Check Adapter with 5% CO Figure 9 Testing the monitor Document No. 896551-2...
Page 341
Datex-Ohmeda Optional N-LDNET Upgrade for Light Monitors, F-LM1 and F-LMP1 Installation Guide All specifications are subject to change without notice. Document No. 896552-2 November 2000 Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O. Box 7550, Madison P.O. Box 900 WI 53707-7550, USA Tel.
Page 343
Table of contents TABLE OF CONTENTS Optional N-LDNET Upgrade for S/5™ Light Monitors, F-LM1 and F-LMP1 TABLE OF CONTENTS TABLE OF FIGURES INTRODUCTION Installation time ........................... 1 Tools needed ............................1 Parts needed ............................1 CONFIGURATION AND TESTS Network and DataCard configuration and test..................2 Network Test ............................
Page 345
Optional N-LDNET Upgrade for Light Monitors, F-LM1 and F-LMP1 INTRODUCTION This guide describes how to install the DataCard & Network upgrade set for Light Monitors, F-LM1 and F-LMP1. The DataCard & Network upgrade set provides a patient trend data collection to Datex-Ohmeda PCMCIA MemCard and Network connection to Datex-Ohmeda Information Centre.
Page 346
Set time & date ( – Set Time and Date) and make user configurations. 2.2 Network Test Check that the monitor connects to the Datex-Ohmeda Network, i.e. the network symbol is displayed on the upper right-hand corner of the screen.
Page 347
Optional N-LDNET Upgrade for Light Monitors, F-LM1 and F-LMP1 891437 892524 892523 617120 617210 617210 Figure 1 Parts NET and Data Data Figure 2 Three different PC/Ethernet boards Document No. 896552-2...
Page 348
Datex-Ohmeda S/5 Light Monitor Figure 3 Data card hardware is installed on top of the recorder housing Ethernet cable with connector ID cable with connector Figure 4 Fasten the Ethernet cable with connector and the ID cable with connector on the metal plate Document No.
Page 349
Optional N-LDNET Upgrade for Light Monitors, F-LM1 and F-LMP1 Figure 5 Lift the rear of the board before pushing the connector in Cover plate Hole Figure 6 Remove the cover plate and install the new one with the hole Document No. 896552-2...
Page 350
Datex-Ohmeda S/5 Light Monitor Data card Figure 7 Only Data card, no menu card allowed Document No. 896552-2...
Page 351
Datex-Ohmeda Optional N-LDATA Upgrade for Light Monitors, F-LM1 and F-LMP1 Installation Guide 0537 Conformity according to the Council Directive 93/42/EEC concerning Medical Devices All specifications are subject to change without notice. Document No. 896553-2 November 2000 Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O.
Page 353
Table of contents TABLE OF CONTENTS Optional N-LDATA Upgrade for S/5™ Light Monitors, F-LM1 and F-LMP1 TABLE OF CONTENTS TABLE OF FIGURES INTRODUCTION Installation time ........................... 1 Tools needed ............................1 Parts needed ............................1 TABLE OF FIGURES Figure 1 Parts..............................
Page 355
Optional N-LDATA Upgrade for Light Monitors, F-LM1 and F-LMP1s INTRODUCTION This guide describes how to install the DataCard upgrade set for Light Monitors, F-LM1 and F-LMP1. The DataCard upgrade set provides patient trend data collection to Datex-Ohmeda PCMCIA MemCard. With the DataCard, the patient trend data collected with the Light Monitor can be transferred to other S/5 monitors.
Page 356
Datex-Ohmeda S/5 Light Monitor 894697 887045 893310 617210 Figure 1 Parts Three different PC/Ethernet boards: NET and Data Data Figure 2 Three Ethernet boards Document No. 896553-2...
Page 357
Optional N-LDATA Upgrade for Light Monitors, F-LM1 and F-LMP1s Data card hardware is installed on top of the recorder housing. Figure 3 Installing Data card hardware Lift the rear of the board before pushing the connector in Figure 4 Installing the board...
Page 358
Datex-Ohmeda S/5 Light Monitor Remove the cover plate and install the new one with the hole Figure 5 Replacing the cover plate Note: Only Data card, no Menu card allowed Figure 6 Data card inserted Document No. 896553-2...
Page 359
Datex-Ohmeda Optional N-LNET Upgrade for Light Monitors, F-LM1 and F-LMP1 Installation Guide 0537 Conformity according to the Council Directive 93/42/EEC concerning Medical Devices All specifications are subject to change without notice. Document No. 896554-2 November 2000 Datex-Ohmeda, Inc. GE Healthcare Finland Oy Helsinki, Finland P.O.
Page 361
Table of contents TABLE OF CONTENTS Optional N-LNET Upgrade for Light Monitors, F-LM1, F-LMP1 S/5™ Light Monitors, F-LM1 and F-LMP1 TABLE OF CONTENTS TABLE OF FIGURES INTRODUCTION Installation time ........................... 1 Tools needed ............................1 Parts needed ............................1 TABLE OF FIGURES Figure 1 Parts..............................
Page 363
Optional N-LNET Upgrade for Light Monitors, F-LM1, F-LMP1 INTRODUCTION This guide describes how to install the Network upgrade set for S/5 Light Monitors, F-LM1 and F- LMP1. The Network upgrade set provides network connection to Datex-Ohmeda Information Centre for central monitoring functions and printing services.
Page 364
Datex-Ohmeda S/5 Light Monitor 894698 892524 892523 617120 617210 617210 Figure 1 Parts Three different PC/Ethernet boards: NET and Data Data Three different PC/Ethernet boards: Figure 2 Document No. 896554-2...
Page 365
Optional N-LNET Upgrade for Light Monitors, F-LM1, F-LMP1 The network hardware is to be installed on top of the recorder housing. Figure 3 Installing Network hardware Fasten the Ethernet cable with connector and the ID cable with connector on the metal plate.
Page 366
Datex-Ohmeda S/5 Light Monitor Lift the rear end of the Ethernet board when pushing the board into the connector. Figure 5 Placing the Ethernet board After completing the upgrade: Figure 6 Upgrade completed Document No. 896554-2...
Page 367
Method of Replacing the Lamp Unit (LCD Backlight)
Page 369
METHOD OF REPLACING THE LAMP UNIT (1) Precautions Please pay attention to the following items while replacing the Lamp Unit. a. Please do not damage the LCD Panel Surface, and do not touch it with bare hands. (Wearing gloves is recommended.) b.
Page 370
Stand the TFT-LCD Module up and push down the latch with that the Lamp Unit fastens. Latch Push down the latch. Picture 3 Pull the cable slowly and remove the Lamp Unit. Picture 4 Picture 5 Remove the other Lamp Unit at the opposite(bottom) side of LCD Module in the same way.
Page 371
Prepare to insert the new Lamp Units. • Open the package and take the new Lamp Units out. • Check the new Lamp Units for dust and foreign matters. Stand the TFT-LCD module and insert the new Lamp Unit. The light guide and reflector sheet should be inside of the Lamp Unit. Please pay attention to insert direction.
Page 372
Fasten the wires to the latch of the plastic frame. Fasten the Pink and white wires to the latch of the plastic frame. Picture 9 Insert the other Lamp Unit to the opposite(bottom) side of the LCD module in the same way. See.
Page 373
WARNING This service manual is available in English only. • If a customer's service provider requires a language other than English, it is (EN) the customer's responsibility to provide translation services. • Do not attempt to service the equipment unless this service manual has been consulted and is understood.
Page 374
HOIATUS Käesolev teenindusjuhend on saadaval ainult inglise keeles. • Kui klienditeeninduse osutaja nõuab juhendit inglise keelest erinevas keeles, (ET) vastutab klient tõlketeenuse osutamise eest. • Ärge üritage seadmeid teenindada enne eelnevalt käesoleva teenindusjuhendiga tutvumist ja sellest aru saamist. • Käesoleva hoiatuse eiramine võib põhjustada teenuseosutaja, operaatori või patsiendi vigastamist elektrilöögi, mehaanilise või muu ohu tagajärjel.
Page 375
警告 このサービスマニュアルは英語版しかありません。 ・ サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業 (JA) はその業者の責任で行うものとさせていただきます。 ・ このサービスマニュアルを熟読し、十分に理解した上で装置のサービスを 行ってください。 ・ この警告に従わない場合、サービスを担当される方、操作員あるいは患者 が、感電や機械的又はその他の危険により負傷する可能性があります。 경고 본 서비스 지침서는 영어로만 이용하실 수 있습니다 . • 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우 , 번역 서비스를 제 (KO) 공하는 것은 고객의 책임입니다 . •...
Page 376
AVISO Este manual técnico só se encontra disponível em inglês. • Se a assistência técnica do cliente solicitar estes manuais noutro idioma, é (PT-PT) da responsabilidade do cliente fornecer os serviços de tradução. • Não tente reparar o equipamento sem ter consultado e compreendido este manual técnico.
Page 377
ЗАСТЕРЕЖЕННЯ Дане керівництво з сервісного обслуговування постачається виключно англійською мовою. (UK) • Якщо сервісний інженер потребує керівництво іншою мовою, користувач забов’язаний забезпечити послуги перекладача. • Не намагайтеся здійснювати технічне обслуговування даного обладнання, якщо ви не читали, або не зрозуміли інформацію, надану в керівництві...
Need help?
Do you have a question about the F-LM1 and is the answer not in the manual?
Questions and answers