9.4.4 To replace
the LCD display
A
1009-0357-000 11/03
1. Disconnect the following cables:
• Membrane switch flex-cable at ZIF (zero insertion force) connector (A)
• Encoder assembly cable (B)
• IRDA Interface cable (C) — remove IRDA Interface board (20)
• Membrane switch flex-cable at ZIF (zero insertion force) connector (D)
• Fan cable (E)
• LCD cable (F)
2. Remove the ten screws (circled) that hold the mounting plate to the front enclosure.
(20)
B
C
3. Remove the mounting plate assembly from the front enclosure.
4. Disconnect the backlight harnesses (F) from the inverter boards.
5. Slide the grommet (G) out of the mounting plate slot (transfer to new LCD).
6. Remove the four screws (circled) that hold the LCD to the mounting plate.
9 Repair Procedures
D
E
G
F
F