To Replace The Backlights - Datex-Ohmeda S/5 Avance Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

S/5 Avance
9.4.5 To replace
the backlights
9-10
7. Lift the left side of the LCD display slightly away from the mounting plate to pull some
of the display ribbon cable (H) to the top side of the plate. Flip the LCD over to the left
of the assembly.
8. Disconnect the display ribbon cable (I).
Display
9. Reassemble in reverse order.
Note: When replacing the LCD, pull the excess ribbon cable to the bottom side of the
plate as you lower the LCD on to the plate. For the backlight harness grommet (G),
ensure that the slit in the grommet faces toward the inside of the keyhole.
The backlight replacement kit includes a backlight assembly (with two backlights) and
two inverters with mounting hardware. To replace the backlight assembly follow the
procedure in Section 9.4.4 to gain access to the assembly. To replace the inverters,
follow the procedure in the next section.
1. Remove the one screw (J) that holds the backlight assembly to the LCD.
2. Slide the backlight assembly to the left to free it from the retaining tabs and then lift
it out of the holder.
Slit
3. Transfer the grommet (G) to the new backlight assembly.
4. Reassemble in reverse order.