9.12.3 Replace
vaporizer manifold
check valve
9-34
1. Set the system switch to Standby.
2. Remove the vaporizers from the vaporizer manifold.
3. Remove the upper rear panel.
4. Disconnect the tubing from the valve block.
5. Remove the valve block.
• To access the left-hand mounting screw (A), remove the right (viewed from front)
side panel (B).
• The right-side extrusion include an access hole (C) for removing the left-hand
mounting screw.
• Note: For early production machines that do not have an access hole in the
extrusion, you must remove the vaporizer manifold to remove the valve block.
C
B
A
11/03 1009-0357-000