S/5 Avance

5.1 Primary Regulators

w w w w WARNING
5-2
Follow the procedure in Section 5.1.1 to gain access to the regulators. Then,
in Section 5.1.2, select the test that is appropriate for the regulator you are
testing.
When testing/adjusting N
system. Use a safe and approved procedure to collect and remove it.
To test or calibrate the primary regulators, you must boot the system with the
PCMCIA Service Application and access the Gas Diagnostics function as
detailed in Section 8.
O regulators, nitrous oxide flows through the
2
11/03 1009-0357-000