The Logging Application; Figure 145 Example Logging Application - Agilent Technologies 8163A Manual

Lightwave multimeter
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Applications
242
All manuals and user guides at all-guides.com

The Logging Application

The Logging data acquisition application logs a series of power
measurements for a number of Power Meter channels, plots the results as
a graph, and generates a statistical analysis of the results.
1
2

Figure 145 Example Logging Application

The most important parameters for the logging application are:
[Samples], this is the number of samples that will be performed during the
logging application, and
[AvgTime], the averaging time is the length of time over which a signal is
averaged. Longer averaging times increase the accuracy and improve the noise
rejection. Longer averaging times also decrease sensitivity.
The total time for a logging application is the number of samples multiplied
by the averaging time.
The Logging application differs from the "The Stability Application" on
page 254 because all power measurements are performed consecutively
without the possibility of any pause between measurements, as shown in
Figure 145 .
All results can be:
• displayed on the screen,
• printed out to hardcopy, or
• saved to disk drive of your Agilent 8164A/B Lightwave Measurement System.
Averaging Time
3
4
5
6
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition

The Logging Application

7
8
9
t
t