16.3.2 Zeroing the Pressure Transducer

To avoid inaccurate pressure readings, the monitor requires a valid zero. Zero the transducer in
accordance with your hospital policy (at least once per day). You must perform a zero:
When you use a new transducer or tubing;
Every time you reconnect the transducer cable to the monitor;
If you think the monitor's pressure readings are not correct.
When using a pressure module, the zero information is stored in the module.
The zeroing procedure is listed as below:
1. Turn off the stopcock to the patient.
2. Vent the transducer to atmospheric pressure, to compensate for the static and atmospheric
pressure exerted on the transducer.
3. In the setup menu for the pressure, select Zero Channel: XX or Zero All . ( XX stands for the
IBP label name). After confirmation, the user can zero the pressure of certain channel or
pressure of all channels. After zeroing, the interface displays the result and last calibration
time.
4. When you see the message Zero Ok , please close the stopcock to atmospheric pressure, and
open the stopcock to the patient.

16.3.3 Troubleshooting the Pressure Zeroing (Taking Art for Example)

The status message lists the probable cause of an unsuccessful calibration.
Message
Art ZERO FAIL
Art SENSOR OFF, FAIL
IN DEMO, FAIL
PRESSURE OVER RANGE,
FAIL
PULSATILE
ZERO FAIL

16.3.4 IBP Calibration

IBP is not user-calibrated. Calibration should be performed by a qualified service professional as
Corrective Action
Make sure that the transducer is not attached to the patient.
Make sure that transducer is not off, and then proceed
zeroing.
Make sure that the monitor is not in DEMO mode. Contact
service technician if necessary.
Make sure that the stopcock is vented to atmosphere. If the
problem persists, please contact service technician.
PRESSURE
Make sure that the transducer is vented to air, not connected
to a patient, and try again.