EDAN iM50 User Manual page 72

Patient monitor
Hide thumbs Also See for iM50:
Table of Contents

Advertisement

Patient Monitor User Manual
Message
NIBP System Failure
NIBP Weak Signal
NIBP
Range
Exceeded
SYS(NIBP)
Overrange
DIA(NIBP)
Overrange
MAP(NIBP)
Overrange
NIBP Loose Cuff
NIBP Interference
NIBP
Leak
Test
Error
(EDAN NIBP module)
NIBP Pressure Low
(Omron NIBP module)
NIBP
Pulse
Abnormal
(Omron NIBP module)
Cause
NIBP is not calibrated.
Cuff is too loose or patient
pulse is too weak.
All of the SYS, DIA and
MAP value are beyond the
measurement range.
SYS
(NIBP)
value
beyond the measurement
range.
DIA
(NIBP)
value
beyond the measurement
range.
MAP
(NIBP)
value
beyond the measurement
range.
Cuff
is
not
wrapped or no cuff is
connected.
Signal noise is too large or
pulse rate is not regular due
to the patient movement.
Fail to deflate normally
during the leak test, so
NIBP leak test cannot be
finished.
Maybe
pre-inflation
pressure can't block the
vessel.
Arrhythmia is serious or
pulse rate is not regular due
to the patient movement.
Alarm Level
High
Low
High
is
High
is
High
is
High
properly
Low
Low
Low
Low
Low
- 60 -
Alarm Information
Action Taken
Contact your service
personnel.
Use other methods to
measure
blood
pressure.
Use other methods to
measure
blood
pressure.
Use other methods to
measure
blood
pressure.
Properly wrap the
cuff.
Make sure that the
patient
under
monitoring
is
motionless.
Test again. If the
problem still exists,
contact your service
personnel.
Check if the cuff
leak
or
cuff
is
properly wrapped.
Make sure that the
patient
under
monitoring
is
motionless.